Assessment of the optical radiation hazard from a home-use intense pulsed light (IPL) source.

Ewan Eadie1, Paul Miller, Teresa Goodman, Harry Moseley.   

Abstract

BACKGROUND AND OBJECTIVES: Intense pulsed light (IPL) systems have evolved and crossed over from the clinic to the home. Studies have shown home-use IPLs to be clinically effective but there has been no published data on ocular safety. It was our aim to measure the spectral and temporal optical radiation output from a home-use IPL and assess the ocular hazard.
MATERIALS AND METHODS: The iPulse Personal is a new home-use IPL hair reduction system. We measured its optical radiation spectral output using a calibrated diode array spectrometer that was traceable to national standards. Pulse duration was determined by measurement with a fast photodiode. The results from these measurements were used to assess the optical radiation hazard to the human eye. Retinal thermal hazard (RTH), blue light hazard (BLH), and infrared radiation hazard to the cornea and lens were assessed in accordance with IEC TR 60825-9 and the International Committee on Non-Ionizing Radiation Protection (ICNIRP) Guidelines on Limits of Exposure to Broad-band Incoherent Optical Radiation, as there are no specific international IPL standards.
RESULTS: Neither the BLH radiance dose nor the infrared radiation hazard to the cornea and lens irradiance exceeded the exposure limit values (ELVs) set by the ICNIRP. The RTH radiance, however, was exceeded at a fluence of 11 J cm(-2) and pulse duration of 16 milliseconds. Following these results the settings on the IPL were adjusted and the RTH was no longer exceeded at a new fluence of 10 J cm(-2) and pulse duration of 26 milliseconds.
CONCLUSIONS: The home-use device that we assessed does not present an optical hazard according to currently available international standards.

Entities:  

Mesh:

Year:  2009        PMID: 19639624     DOI: 10.1002/lsm.20801

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  2 in total

Review 1.  Are home-use intense pulsed light (IPL) devices safe?

Authors:  Godfrey Town; Caerwyn Ash
Journal:  Lasers Med Sci       Date:  2010-07-13       Impact factor: 3.161

2.  Effects of intense pulsed light on tissue vascularity and wound healing: a study with mouse island skin flap model.

Authors:  Trinh Cao Minh; Do Xuan Hai; Pham Thi Ngoc
Journal:  Plast Surg Int       Date:  2015-02-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.