Literature DB >> 19635380

On-line solid phase extraction LC-MS/MS analysis of pharmaceutical indicators in water: a green alternative to conventional methods.

Rebecca A Trenholm1, Brett J Vanderford, Shane A Snyder.   

Abstract

A method using automated on-line solid phase extraction (SPE) directly coupled to liquid chromatography/tandem mass spectrometry (LC-MS/MS) has been developed for the analysis of six pharmaceuticals by isotope dilution. These selected pharmaceuticals were chosen as representative indicator compounds and were used to evaluate the performance of the on-line SPE method in four distinct water matrices. Method reporting limits (MRLs) ranged from 10 to 25 ng/L, based on a 1 mL extraction volume. Matrix spike recoveries ranged from 88 to 118% for all matrices investigated, including finished drinking water, surface water, wastewater effluent and septic tank influent. Precision tests were performed at 50 and 1000 ng/L with relative standard deviations (RSDs) between 1.3 and 5.7%. A variety of samples were also extracted using a traditional off-line automated SPE method for comparison. Results for both extraction methods were in good agreement; however, on-line SPE used approximately 98% less solvent and less time. On-line SPE coupled to LC-MS/MS analysis for selected indicators offers an alternative, more environmentally friendly, method for pharmaceutical analysis in water by saving time and costs while reducing hazardous waste and potential environmental pollution as compared with off-line SPE methods.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19635380     DOI: 10.1016/j.talanta.2009.06.006

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

1.  Simultaneous quantitative monitoring of four indicator contaminants of emerging concern (CEC) in different water sources of Central India using SPE/LC-(ESI)MS-MS.

Authors:  Roshan Appa; V A Mhaisalkar; Amit Bafana; S Saravana Devi; Kannan Krishnamurthi; Tapan Chakrabarti; Pravin K Naoghare
Journal:  Environ Monit Assess       Date:  2018-07-25       Impact factor: 2.513

2.  Automatic analytical approach for the determination of 12 illicit drugs and nicotine metabolites in wastewater using on-line SPE-UHPLC-MS/MS.

Authors:  Jingyuan Wang; Likai Qi; Chenzhi Hou; Tingting Zhang; Mengyi Chen; Haitao Meng; Mengxiang Su; Hui Xu; Zhendong Hua; Youmei Wang; Bin Di
Journal:  J Pharm Anal       Date:  2021-01-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.