Literature DB >> 19634120

A sensitive liquid chromatography-electrospray ionization-mass spectrometry method for the simultaneous determination of pentoxyverine citrate and guaifenesin in human plasma---application to pharmacokinetic and bioequivalence studies.

Jinhua Wen1, Hong Zhang, Chunhua Xia, Xiao Hu, Wenwei Xu, Xiaohua Cheng, Jun Gao, Yuqing Xiong.   

Abstract

A sensitive and specific liquid chromatography-electrospray ionization-mass spectrometry method for the identification and quantification of pentoxyverine citrate and guaifenesin in human plasma has been developed. After extraction from plasma samples by ethyl acetate, the internal standard and analytes were separated by high-performance liquid chromatographic on a Shim-pack VP-ODS C(18) column (150 x 2.0 mm) using a mobile phase consisting of A (methanol) and B (0.4% glacial acetic acid and 4 mmol/L ammonium acetate) (A:B, 43 : 57). Analysis was performed on a Shimadzu LC/MS-2010A in selected ion monitoring mode with a positive electrospray ionization interface. The method was linear in the concentration range of 1.0-640.0 ng/mL for pentoxyverine citrate and 0.025-6.4 microg/mL for guaifenesin. The inter- and intra- precision were all within 12% and accuracy ranged from 85 to 115%.The lower limits of quantification were 1.0 ng/mL for pentoxyverine citrate and 25.0 ng/mL for guaifenesin. The extraction recovery was on average 81.95% for pentoxyverine citrate and 89.03% for guaifenesin. This is the first assay method reported for the simultaneous determination of pentoxyverine citrate and guaifenesin in plasma using one chromatographic run. Copyright (c) 2009 John Wiley & Sons, Ltd.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 19634120     DOI: 10.1002/bmc.1298

Source DB:  PubMed          Journal:  Biomed Chromatogr        ISSN: 0269-3879            Impact factor:   1.902


  1 in total

1.  Eco-friendly UPLC-MS/MS analysis of possible add-on therapy for COVID-19 in human plasma: Insights of greenness assessment.

Authors:  Omnia A El-Naem; Sarah S Saleh
Journal:  Microchem J       Date:  2021-04-02       Impact factor: 4.821

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.