Literature DB >> 19622897

Validation of arteriograph - a new oscillometric device to measure arterial stiffness in patients on maintenance hemodialysis.

János Nemcsik1, József Egresits, Taha El Hadj Othmane, Bertalan Csaba Fekete, Erzsébet Fodor, Tamás Szabó, Zoltán Járai, Csaba Jekkel, István Kiss, András Tislér.   

Abstract

BACKGROUND: Measuring arterial stiffness (augmentation index (AI), aortic pulse wave velocity (PWV)) in hemodialysis (HD) patients has prognostic significance. To assess its validity, the new oscillometric Arteriograph device (AI(A), PWV(A)) was compared to the validated PulsePen tonometer (AI(P), PWV(P)).
METHODS: AI and PWV were measured in 98 patients with both devices before HD. Validity was evaluated by Pearson's correlation, Bland-Altman analysis, and by assessing the prognostic value of AI and PWV to predict cardiovascular (CV) mortality over 29 months.
RESULTS: Correlation between AI(P) and AI(A) was significant (R = 0.527, p < 0.001). The mean difference of AI values obtained by the two devices was -20.6%, and 30% of the paired AI differences fall outside the +/-1 SD boundary of the mean between-device difference. There was no significant correlation between the PWV(P) and PWV(A) readings (R = 0.173, p = 0.097). The average difference of PWV values by the two devices was -1.2 m/s, and 20.6% of the paired PWV differences fall outside the +/-1 SD boundary. In survival analyses, only PWV(P) but not PWV(A) was significantly related to CV mortality.
CONCLUSION: Lack of correlation between PWV(P) and PWV(A) and lack of prognostic significance of PWV(A) suggest limited validity of Arteriograph to determine PWV in patients on HD. Copyright (c) 2009 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2009        PMID: 19622897     DOI: 10.1159/000228935

Source DB:  PubMed          Journal:  Kidney Blood Press Res        ISSN: 1420-4096            Impact factor:   2.687


  7 in total

1.  Validation of Central and Peripheral Non-Invasive Hemodynamic Variables Using an Oscillometric Method.

Authors:  Darío Echeverri; Alejandro Pizano; Jaime Cabrales; Karen Moreno
Journal:  High Blood Press Cardiovasc Prev       Date:  2017-10-28

Review 2.  Twenty-Four-Hour Ambulatory Pulse Wave Analysis in Hypertension Management: Current Evidence and Perspectives.

Authors:  Stefano Omboni; Igor N Posokhov; Yulia V Kotovskaya; Athanase D Protogerou; Jacques Blacher
Journal:  Curr Hypertens Rep       Date:  2016-10       Impact factor: 5.369

3.  Aortic pulse wave velocity in haemodialysis patients is associated with the prescription of active vitamin D analogues.

Authors:  Evangelina Charitaki; Andrew Davenport
Journal:  J Nephrol       Date:  2014-01-29       Impact factor: 3.902

Review 4.  Ambulatory arterial stiffness in chronic kidney disease: a methodological review.

Authors:  Andrea László; György Reusz; János Nemcsik
Journal:  Hypertens Res       Date:  2015-12-03       Impact factor: 3.872

5.  Catheter-based renal sympathetic denervation improves central hemodynamics and arterial stiffness: a pilot study.

Authors:  Kai Mortensen; Klaas Franzen; Frank Himmel; Frank Bode; Heribert Schunkert; Joachim Weil; Michael Reppel
Journal:  J Clin Hypertens (Greenwich)       Date:  2012-09-11       Impact factor: 3.738

6.  Can arterial wave augmentation in young adults help account for variability of cardiovascular risk in different British ethnic groups?

Authors:  Luca Faconti; Maria J Silva; Oarabile R Molaodi; Zinat E Enayat; Aidan Cassidy; Alexis Karamanos; Elisa Nanino; Ursula M Read; Philippa Dall; Ben Stansfield; Seeromanie Harding; Kennedy J Cruickshank
Journal:  J Hypertens       Date:  2016-11       Impact factor: 4.844

7.  Feasibility and characteristics of arterial stiffness measurement in preschool children.

Authors:  Maaike M Vanderschuren; Cuno Spm Uiterwaal; Cornelis K van der Ent; Jacobien B Eising
Journal:  Eur J Prev Cardiol       Date:  2017-07-21       Impact factor: 7.804

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.