A prototype high-purity germanium detector system with fast photon-counting circuitry for medical imaging.

B H Hasegawa1, B Stebler, B K Rutt, A Martinez, E L Gingold, C S Barker, K G Faulkner, C E Cann, D P Boyd.   

Abstract

A data-acquisition system designed for x-ray medical imaging utilizes a segmented high-purity germanium (HPGe) detector array with 2-mm wide and 6-mm thick elements. The detectors are contained within a liquid-nitrogen cryostat designed to minimize heat losses. The 50-ns pulse-shaping time of the preamplifier electronics is selected as the shortest time constant compatible with the 50-ns charge collection time of the detector. This provides the detection system with the fastest count-rate capabilities and immunity from microphonics, with moderate energy resolution performance. A theoretical analysis of the preamplifier electronics shows that its noise performance is limited primarily by its input capacitance, and is independent of detector leakage current up to approximately 100 nA. The system experimentally demonstrates count rates exceeding 1 million counts per second per element with an energy resolution of 7 keV for the 60-keV gamma ray photon from 241Am. The results demonstrate the performance of a data acquisition system utilizing HPGe detector systems which would be suitable for dual-energy imaging as well as systems offering simultaneous x-ray transmission and radionuclide emission imaging.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1961152     DOI: 10.1118/1.596606

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Gallium SPECT/CT in lymphoma: the ups and downs of functional imaging.

Authors:  Rachel Bar-Shalom
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-11       Impact factor: 9.236

Review 2.  Technological development and advances in single-photon emission computed tomography/computed tomography.

Authors:  Youngho Seo; Carina Mari; Bruce H Hasegawa
Journal:  Semin Nucl Med       Date:  2008-05       Impact factor: 4.446

3.  Development of a Germanium Small-Animal SPECT System.

Authors:  Lindsay C Johnson; Oleg Ovchinnikov; Sepideh Shokouhi; Todd E Peterson
Journal:  IEEE Trans Nucl Sci       Date:  2015-10-09       Impact factor: 1.679

4.  Characterization of a high-purity germanium detector for small-animal SPECT.

Authors:  Lindsay C Johnson; Desmond L Campbell; Ethan L Hull; Todd E Peterson
Journal:  Phys Med Biol       Date:  2011-08-18       Impact factor: 3.609

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.