Abstract
Exemplars of current commercial breath-alcohol simulators were studied to ascertain their suitability for control tests and as calibrators in breath-alcohol analysis. Temperature, effluent volumes and pressures, effluent-alcohol concentration, and alcohol depletion were measured for simulators coupled with various current alcohol analyzers. Effluent recirculation was found to extend the number of acceptable control test cycles by 3.2 x over nonrecirculation. We found substantial improvements in current commercial simulators and in their performance over those evaluated in 1979.Entities:
Mesh:
Substances:
Year: 1991 PMID: 1960980 DOI: 10.1093/jat/15.5.272