| Literature DB >> 19609034 |
Akira Kotani1, Yanan Yuan, Bin Yang, Yuzuru Hayashi, Rieko Matsuda, Fumiyo Kusu.
Abstract
The optimum conditions of the mobile phase for HPLC with electrochemical detection (HPLC-ECD) were selected from among solvents of different grades with the standard deviation (SD) of area measurements based on FUMI (function of mutual information) theory as a criterion. In HPLC-ECD for determining baicalin and baicalein using a phosphoric acid-methanol-water mixture (0.5:60:40, v/v/v) as a mobile phase, the effects of three grades of phosphoric acid, of five grades of methanol, and of seven grades of water on the measurement precision were examined. The mixture, if characterized by the minimum SD of measurements, was selected as the optimal mobile phase. This selection method saves considerable amounts of chemicals and experimental time, and would be a useful exploration technique for the routine check and troubleshooting of HPLC-ECD.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19609034 DOI: 10.2116/analsci.25.925