[Measurement of residual effect in the amorphous selenium flat panel detector system].

Kojiro Nishijima1, Katsuhiko Ueda, Shohei Kudomi, Taizo Sanada.   

Abstract

The purpose of this study was to evaluate the residual effect generated by the amorphous selenium flat panel detector system (a-Se FPD). A residual effect occurs as a result of the addition of delayed electrons by previous X-ray irradiation joining the signal and change in detector sensitivity caused by hole-electron recombination or trapped electrons in a-Se. To evaluate the effect of previous radiation exposure, we irradiated a-Se FPD that were half-shielded by a 3 mm thick lead plate. A residual effect was generated in irradiated areas, with the unirradiated areas serving as reference points. Next, we removed the lead plate and took a new image using uniform irradiation. The difference in pixel value between irradiated and nonirradiated areas was measured using a variety of time intervals between each exposure. Through a comparison of pixel values from images taken over various time intervals, we discovered our system needs 20 hours to return to a normal state and become capable of producing a residual-free image.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19602803     DOI: 10.6009/jjrt.65.782

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  1 in total

1.  Evaluation of image lag in a flat-panel, detector-equipped cardiovascular X-ray machine using a newly developed dynamic phantom.

Authors:  Hisaya Sato; Keisuke Kondo; Kyoichi Kato; Yasuo Nakazawa
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.