Comparison of tilted versus nontilted implant-supported prosthetic designs for the restoration of the edentuous mandible: a biomechanical study.

Chiara M Bellini1, Davide Romeo, Fabio Galbusera, Silvio Taschieri, Manuela T Raimondi, Antonios Zampelis, Luca Francetti.   

Abstract

PURPOSE: The purpose of the study was to evaluate, using finite element analysis, the stress patterns induced in cortical bone by three distinct implant-supported prosthetic designs.
MATERIALS AND METHODS: The first two models consisted of a prosthesis supported by four implants, the distal two of which were tilted, with different cantilever lengths (5 mm and 15 mm). The third design consisted of a prosthesis supported by five conventionally placed implants and a 15-mm cantilever.
RESULTS: In the tilted model with 5-mm cantilever and in the nontilted model, the maximum value of compressive stress (-18 MPa) was found near the cervical area of the distal implant. Higher values for compressive stress were predicted near the cervical area of the distal implant in the tilted model with a 15-mm cantilever, as compared to the tilted model with the 5-mm cantilever. For the tilted model with the 5-mm cantilever, peak values of tensile stress were predicted near the cervical area of both the distal (1.25 MPa) and the mesial implants (2.5 MPa). For the nontilted model, the peak value was found near the cervical area of the in-between implant (5 MPa). For the tilted model with 15-mm cantilever, tensile stress values were higher than in the tilted model with 5-mm cantilever.
CONCLUSIONS: No significant difference in stress patterns between the tilted 5-mm and the nontilted 15-mm configuration was predicted. The tilted configuration with a 15-mm cantilever was found to induce higher stress values than the tilted configuration with a 5-mm cantilever.

Mesh:

Substances:

Year:  2009        PMID: 19587875

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  12 in total

Review 1.  Prosthodontic perspective to all-on-4® concept for dental implants.

Authors:  M Taruna; B Chittaranjan; N Sudheer; Suchita Tella; Md Abusaad
Journal:  J Clin Diagn Res       Date:  2014-10-20

Review 2.  Oral health-related quality of life and satisfaction in edentulous patients rehabilitated with implant-supported full dentures all-on-four concept: a systematic review.

Authors:  Gabriela Sumie Yaguinuma Gonçalves; Keith Murieli Ferreira de Magalhães; Eduardo Passos Rocha; Paulo Henrique Dos Santos; Wirley Gonçalves Assunção
Journal:  Clin Oral Investig       Date:  2021-10-13       Impact factor: 3.573

3.  Evaluation of immediately loaded mandibular four vertical versus tilted posterior implants supporting fixed detachable restorations without versus with posterior cantilevers.

Authors:  Latifa Abdallah Mohamed; Mohamed Moataz Khamis; Ahlam Mostafa El-Sharkawy; Rania Abdelaziz Fahmy
Journal:  Oral Maxillofac Surg       Date:  2021-08-29

4.  Does matching relation exist between the length and the tilting angle of terminal implants in the all-on-four protocol? stress distributions by 3D finite element analysis.

Authors:  Xiaomei Li; Zhizhong Cao; Xiaoqian Qiu; Zhen Tang; Lulu Gong; Dalin Wang
Journal:  J Adv Prosthodont       Date:  2015-06-23       Impact factor: 1.904

5.  Angulated Implants for Fabrication of Implant Supported Fixed Partial Denture in the Maxilla.

Authors:  Nicholas Egbert; Swati Ahuja; Audrey Selecman; Russell Wicks
Journal:  J Dent (Shiraz)       Date:  2017-12

6.  Peri-Implant Behavior of Sloped Shoulder Dental Implants Used for All-On-Four Protocols: An Histomorphometric Analysis in Dogs.

Authors:  Jose Luis Calvo Guirado; Aldo Fabian Lucero-Sánchez; Ana Boquete Castro; Marcus Abboud; Sergio Gehrke; Manuel Fernández Dominguez; Rafael Arcesio Delgado Ruiz
Journal:  Materials (Basel)       Date:  2018-01-12       Impact factor: 3.623

7.  Effect of the number of abutments on biomechanics of Branemark prosthesis with straight and tilted distal implants.

Authors:  Marcos Michelon Naconecy; Tomás Geremia; André Cervieri; Eduardo Rolim Teixeira; Rosemary Sadami Shinkai
Journal:  J Appl Oral Sci       Date:  2010 Mar-Apr       Impact factor: 2.698

8.  A finite element analysis to study the stress distribution on distal implants in an all-on-four situation in atrophic maxilla as affected by the tilt of the implants and varying cantilever lengths.

Authors:  Anju Kumari; Puja Malhotra; Shefali Phogat; Bhupender Yadav; Jaiveer Yadav; Sumit Singh Phukela
Journal:  J Indian Prosthodont Soc       Date:  2020-10-08

9.  Evaluation of stress distribution of implant-retained mandibular overdenture with different vertical restorative spaces: A finite element analysis.

Authors:  Behnaz Ebadian; Mahmoud Farzin; Saeid Talebi; Niloufar Khodaeian
Journal:  Dent Res J (Isfahan)       Date:  2012-11

10.  Influence of Framework Material and Posterior Implant Angulation in Full-Arch All-on-4 Implant-Supported Prosthesis Stress Concentration.

Authors:  João Paulo Mendes Tribst; Dayana Campanelli de Morais; Jefferson David Melo de Matos; Guilherme da Rocha Scalzer Lopes; Amanda Maria de Oliveira Dal Piva; Alexandre Luiz Souto Borges; Marco Antonio Bottino; Antonio Lanzotti; Massimo Martorelli; Pietro Ausiello
Journal:  Dent J (Basel)       Date:  2022-01-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.