Marian A C Piqueur1, Walter A Verstrepen, Peggy Bruynseels, An H Mertens.
Abstract
We describe an improvement of an earlier reported real-time RT-PCR assay for the detection of enterovirus RNA, based on the 5' exonuclease digestion of a dual-labeled fluorogenic probe by Taq DNA polymerase. A different extraction method, real-time RT-PCR instrument and primer set were evaluated. Our data show that the optimized assay yields a higher sensitivity and reproducibility and resulted in a significant reduced hands-on time per sample.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19583870 PMCID: PMC2716329 DOI: 10.1186/1743-422X-6-95
Source DB: PubMed Journal: Virol J ISSN: 1743-422X Impact factor: 4.099
Compared primer-probe sets
| Primer-probe set 1 (earlier reported) | Primer-probe set 2 | |
| Forward primer | 5'-CCCTGAATGCGGCTAATCC-3' | 5' CCGGCCCCTGAATGC-3' |
| Reverse primer | 5'-ATTGTCACCATAAGCAGCCA-3' | 5' CACCGGATGGCCAATCCA-3' |
| Probe | FAM 5' AACCGACTACTTTGGGTGTCCGTGTTTC-3' TAMRA | |
Mean Ct/Cp values and %CV values for the undiluted OptiQual Enterovirus Run Control in different real-time RT-PCR settings
| Extraction Method | Platform | Mean Ct/Cp | |
| Primer-probe set 1 | Primer-probe set 2 | ||
| QiaAmp | 7700 SDS | 38.47 ± 3.94 | 37.74 ± 1.35 |
| LC 480 | 33.69 ± 2.95 | 34.48 ± 1.67 | |
| EasyMag | 7700 SDS | 38.23 ± 3.88 | 36.29 ± 0.91 |
| LC 480 | 32.81 ± 0.25 | 32.78 ± 0.20 | |
Limit of detection (EVU/ml) for different RT-PCR settings obtained with OptiQual Enterovirus Run Control
| Extraction Method | Platform | LOD (EVU/ml) | |
| primer-probe set 1 | primer-probe set 2 | ||
| QiaAmp | 7700 SDS | 119.5 | 111.8 |
| LC 480 | 54.2 | 11.1 | |
| EasyMag | 7700 SDS | 43.4 | 17.5 |
| LC 480 | 13.5 | 4.5 | |