| Literature DB >> 19579263 |
Tobias Klaassen1, Simon Szwandt, James T Kapron, Axel Roemer.
Abstract
The analysis of peptides presents serious challenges for bioanalytical scientists including low total ion current and non-selective fragmentation during tandem mass spectrometry (MS/MS). During method validation of a peptide in rat serum matrix some interferences could not be easily removed and thus prevented accurate and precise measurement. These problems associated with peptide quantitation were resolved by using FAIMS (high-Field Asymmetric waveform Ion Mobility Spectrometry). This selectivity-enhancing technique filters out matrix interferences, and the resulting pseudo-selected reaction monitoring (pseudo-SRM) chromatograms were nearly free from interferences. Control blank matrix samples contained an acceptable level of interference (only 7% signal as compared to the lower level of quantitation). Chromatographic peaks were easily, accurately and precisely integrated resulting in a validated liquid chromatography (LC)/FAIMS-MS/MS method for the analysis of a peptide drug in rat serum according to United States Food and Drug Administration (US FDA) bioanalytical guidelines. These results confirm that new selectivity-enhancing technologies aid the pharmaceutical industry in reliably producing acceptable pharmacokinetic data. Copyright (c) 2009 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19579263 DOI: 10.1002/rcm.4147
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419