| Literature DB >> 19575109 |
Josilene Luciene Duarte1, Kátia de Freitas Alvarenga, Marcos Roberto Banhara, Ana Dolores Passarelli de Melo, Roberta Moreno Sás, Orozimbo Alves Costa Filho.
Abstract
UNLABELLED: The P300 is and auditory Evoked Potential, called endogenous potential because it reflects the functional use the individual makes of the auditory stimulus, being highly dependent on cognitive skills; among them we list attention and auditory discrimination. It is a procedure of objective evaluation; however, one that depends on the examiner's experience to detect wave peaks, and it is important to use recording methods that facilitate the response presence analysis and result interpretation. AIM: To analyze the P300 Long Latency Auditory Evoked Potential obtained through the use of two active electrodes positioned on Fz and Cz.Entities:
Mesh:
Year: 2009 PMID: 19575109 PMCID: PMC9450669 DOI: 10.1016/s1808-8694(15)30783-7
Source DB: PubMed Journal: Braz J Otorhinolaryngol ISSN: 1808-8686
Parameters used in the study of the P300 Long Latency Evoked Auditory Potential.
| Parameters used in the study of the p300 long latency evoked auditory potential | |
| Stimulus type | Tone burst (20% raro e 80 % freqüente) |
| Tone burst (20% rare and 80 % frequent) | 70 Dbna |
| Stimulus frequency | 2000 Hz (rare); 1000 hz (frequent) |
| Stimulus intensity | 70 Dbhl |
| Stimulus occurence speed | 1 Stimulus per second |
| Electrode types | Ekg/agcl with gel |
| Electrode positioning | Fz and cz (active); a1 and a2 (reference) |
| Pre-amplifier | Channels 1 and 2 - input 1 active; input 2 reference (jumper) |
| Impedance | ≤ 5 KΩ(individual); ≤ 2 kΩ (between electrodes) |
| Band pass filter | 1 To 25 hz |
| Transducer | 3a insertion phones |
Figure 1Illustration of the electrodes placement on a patient's skull according to the international 10–20 system and the cable connections on the pre-amplifier cables of the Auditory Evoked Potentials recording system.
Figure 2Records of the N2 and P3 components from the P300 Long Latency Evoked Auditory Potential, simultaneously captured by the electrodes positioned on Fz and Cz. Markings of the N2 negative and P3 positive peaks.
Descriptive analysis (mean, standard deviation, maximum and minimum values) and comparison between males and females (paired t student test) for the P3 and N2 component latency and P3 component amplitude (amp).
| P300 LONG LATENCY EVOKED AUDITORY POTENTIAL | ||||||
|---|---|---|---|---|---|---|
| Fz | Cz | Cz | ||||
| N2 (ms) | P3 (ms) | P3 amp (μv) | N2 (ms) | P3 (ms) | P3 amp (μv) | |
| Mean ± sd | 230±31 | 339±20 | 1.81±1.06 | 228±31 | 341±23 | 2,12±1,07 |
| Minimum | 183 | 301 | 0.1 | 183 | 301 | 0,3 |
| Maximum | 275 | 369 | 4.3 | 273 | 371 | 3,9 |
| Male x female | 0.95 | 0.92 | 0.17 | 0.94 | 0.69 | 0,14 |
P ≤ 0.05 statistically significant
Pearson's correlation coefficient to compare the N2 and P3 components' latency (ms) and P3 component amplitude (μV), measured in Fz and Cz.
| P300 LONG LATENCY EVOKED AUDITORY POTENTIAL | |||
|---|---|---|---|
| Electrode positioning | Latency n2 | Latêncy p3 | Amplitude p3 |
| Fz x cz | 0.676* | 0.940* | 0.687* |
* The correlations were statistically significant (P < 0.01).
Correlation between the individuals' age by means of the Pearson's correlation test, with latency (ms) of components N2 and P3 and P3 component amplitude (amp-μV), measured in Fz and Cz.
| P300 LONG LATENCY EVOKED AUDITORY POTENTIAL | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Fz | Cz | Cz | ||||||||||
| N2 | P3 | P3 amp | N2 | P3 | P3 amp | |||||||
| Correlation coefficient | ||||||||||||
| 0.02 | 0.89 | 0.29 | 0.09 | -0.19 | 0.28 | 0.03 | 0.86 | 0.19 | 0.28 | -0,08 | 0,64 | |
p≤ 0.05 - statistically significant