Literature DB >> 19563969

Vascular prosthesis rupture caused by contact with rib stump after thoracic aorta replacement.

Hiroshi Yamamoto1, Fumio Yamamoto, Kazuyuki Ishibashi, Yasuharu Noishiki.   

Abstract

A 42-year-old male with Marfan syndrome, who had undergone aortic root and total arch replacement for type-A acute aortic dissection at the age of 40, underwent descending aorta replacement with woven Dacron (Vascutek Ltd., Renfrewshire, Scotland) because of pseudoaneurysm at the site of the distal anastomosis and an enlarged pseudolumen of the dissecting descending aorta. The fourth and eighth ribs were cut at their anterior and posterior sites to allow wide exposure of the entire descending aorta. Postoperative computed tomographic scanning showed that the vascular prosthesis posteriorly contacted the eighth rib stump. On the postoperative day 25, the patient collapsed and developed severe hypotension. Emergency thoracotomy revealed a 6 mm in diameter hole on the posterior side of the vascular prosthesis. One day later, the patient died of cardiac dysfunction resulting from sustained hypotension. Electron microscopic examination of the vascular prosthesis showed that the hole was caused by frayed fabric and disrupted polyester fibers. Our experience warns that a woven polyester vascular prosthesis could rupture within 3 weeks of contacting a rib stump.

Entities:  

Mesh:

Year:  2009        PMID: 19563969     DOI: 10.1016/j.jvs.2009.01.064

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  2 in total

1.  Clinical analysis of non-anastomotic aneurysms of implanted prosthetic grafts.

Authors:  Satoshi Yamamoto; Katsuyuki Hoshina; Hideo Kimura; Hiroyuki Okamoto; Kunihiro Shigematsu; Tetsuro Miyata; Toshiaki Watanabe
Journal:  Surg Today       Date:  2014-04-03       Impact factor: 2.549

2.  A Rare Case of Dacron Graft Rupture due to Friction against a Rib.

Authors:  Katsuhiko Matsuyama; Masahiko Kuinose; Nobusato Koizumi; Toru Iwahashi; Kayo Toguchi; Hitoshi Ogino
Journal:  Thorac Cardiovasc Surg Rep       Date:  2014-07-16
  2 in total