Literature DB >> 19560939

Penetration of Günther Tulip filter struts through an introducer sheath: case report and safety concerns.

Sohail G Contractor1, Alex Merkulov, Waseem Bhatti, Michael Lee, Kim Gardner.   

Abstract

The authors present a patient who had a Günther-Tulip inferior vena cava filter placed under fluoroscopic guidance. The filter struts were seen to penetrate through the introducer sheath at deployment. This was believed to be secondary to a kink in the sheath at the site of venous entry and was due to the patient's nuchal obesity as well as his inability to turn his head to the opposite side. The introducer sheath was then placed through a long reinforced metal sheath through which the filter was then placed without complication.

Entities:  

Mesh:

Year:  2009        PMID: 19560939     DOI: 10.1016/j.jvir.2009.04.068

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  1 in total

1.  Penetration of inferior vena cava filter strut through an introducer sheath inserted from the left internal jugular vein: An anatomical pitfall.

Authors:  Satoshi Yoshimura; Seiji Koga; Tomoo Nakata; Satoshi Ikeda; Hiroaki Kawano; Koji Maemura
Journal:  J Cardiol Cases       Date:  2016-12-12
  1 in total