Literature DB >> 19559206

Nontraditional surgical approaches for implantation of pacemaker and cardioverter defibrillator systems in patients with limited venous access.

Dawn E Jaroszewski1, Gregory T Altemose, Luis R Scott, Komandoor Srivasthan, Patrick A Devaleria, Jesse Lackey, Francisco A Arabia.   

Abstract

BACKGROUND: Indications for placement of implantable cardioverter-defibrillators (ICD) and pacemakers have expanded, and traditional transvenous implantation may not be feasible in patients with aberrant anatomy or venous obstruction. In these settings, successful lead placement has required innovative surgical approaches. A case series of successful placement of these systems in challenging patients is presented.
METHODS: A 2-year retrospective study of patients undergoing placement of minimally invasive epicardial pacing leads or ICD coils was performed.
RESULTS: Eleven patients underwent minimally invasive surgical placement of leads or coils. None were converted to open sternotomy. One required extension to minianterior thoracotomy. Causes of intravenous placement failure included aberrant anatomy with failure to access coronary sinus in 9 and venous occlusion in 2. Four patients had previous operations through a median sternotomy. Procedures included left video-assisted thoracoscopic (VATS) placement of a left ventricular epicardial lead in 8, left VATS conversion to minianterior thoracotomy left ventricular epicardial lead placement in 1, left VATS placement of ICD coil in 1, subxiphoid placement of a right ventricular epicardial lead in 1, subxiphoid ICD coil in 2, and subcutaneous ICD coil placement in 3. Mean hospitalization was 4.6 days. Postoperative hypotension and pulmonary edema occurred in 27% of patients. No patients died.
CONCLUSIONS: Conventional transvenous lead implantation may be difficult or impossible in some patients with aberrant or occluded venous access. Novel surgical approaches with the use of minimally invasive procedures can establish optimally functional pacing and ICD systems without sternotomy and low associated morbidity.

Entities:  

Mesh:

Year:  2009        PMID: 19559206     DOI: 10.1016/j.athoracsur.2009.04.006

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  12 in total

1.  Erdheim-Chester disease: The role of video-assisted thoracoscopic surgery in diagnosing and treating cardiac involvement.

Authors:  Aoife Egan; Dan Sorajja; Dawn Jaroszewski; Farouk Mookadam
Journal:  Int J Surg Case Rep       Date:  2011-12-08

2.  Chronic performance of subxiphoid minimally invasive pericardial Model 20066 pacemaker lead insertion in an infant animal model.

Authors:  Bradley C Clark; Rohan Kumthekar; Paige Mass; Justin D Opfermann; Charles I Berul
Journal:  J Interv Card Electrophysiol       Date:  2019-10-15       Impact factor: 1.900

3.  Permanent pacing in patients without upper limb venous access: a review of current techniques.

Authors:  Swee-Chong Seow; Toon-Wei Lim; Devinder Singh; Wee-Tiong Yeo; Pipin Kojodjojo
Journal:  Heart Asia       Date:  2014-11-27

Review 4.  Clinical experience with subcutaneous implantable cardioverter-defibrillators.

Authors:  Geoffrey F Lewis; Michael R Gold
Journal:  Nat Rev Cardiol       Date:  2015-04-21       Impact factor: 32.419

Review 5.  Localized targeting of biomaterials following myocardial infarction: a foundation to build on.

Authors:  James A Shuman; Jonathan R Zurcher; Ashley A Sapp; Jason A Burdick; Robert C Gorman; Joseph H Gorman; Edie C Goldsmith; Francis G Spinale
Journal:  Trends Cardiovasc Med       Date:  2013-06-06       Impact factor: 6.677

6.  Nontraditional placement of an implantable cardioverter-defibrillator in a heterotaxy patient after the completion of total cavopulmonary connection.

Authors:  Toru Tsukada; Yuji Hiramatsu; Shinya Kanemoto; Lisheng Lin; Miho Takahashi-Igari; Hitoshi Horigome; Shonosuke Matsushita; Yuzuru Sakakibara
Journal:  J Artif Organs       Date:  2013-06-01       Impact factor: 1.731

7.  Alternate method for endocardial pacemaker lead implantation: A hybrid mini-thoracotomy approach.

Authors:  Viveka Kumar; Pradipta Kumar Nayak; Mitendra Singh Yadav; Sangeeta Dhir; Vanita Arora; Vivek Kumar
Journal:  Indian Pacing Electrophysiol J       Date:  2021-01-22

8.  Wire countertraction for sheath placement through stenotic and tortuous veins: The "body flossing" technique.

Authors:  Jeffrey S Arkles; Prakash Goutham Suryanarayana; Mouhannad Sadek; Joshua M Cooper; David S Frankel; Fermin C Garcia; Jay Giri; Robert D Schaller
Journal:  Heart Rhythm O2       Date:  2020-04-27

9.  Minimally invasive epicardial left ventricular lead placement in a case of massive pleural adhesion.

Authors:  Hans-Stefan Hofmann; Michael Ried; Zsolt Sziklavari
Journal:  J Cardiothorac Surg       Date:  2014-04-10       Impact factor: 1.637

10.  Contralateral transvenous left ventricular lead placement of implantable devices with pre-sternal tunnelling in chronically obstructed subclavian veins.

Authors:  Praveen P Sadarmin; Rajesh K Chelliah; Jonathan Timperley
Journal:  Indian Pacing Electrophysiol J       Date:  2015-07-29
View more