Proof of concept: hemodynamic response to long-term partial ventricular support with the synergy pocket micro-pump.

Bart Meyns1, Stefan Klotz, Andre Simon, Walter Droogne, Filip Rega, Bartley Griffith, Robert Dowling, Mark J Zucker, Daniel Burkhoff.   

Abstract

OBJECTIVES: The purpose of this study was to test the hemodynamic effects of partial ventricular support in patients with advanced heart failure.
BACKGROUND: The use of current left ventricular assist devices (VADs) that provide full circulatory support is restricted to critically ill patients because of associated risks. Smaller, less-invasive devices could expand VAD use to a larger pool of less-sick patients but would pump less blood, providing only partial support.
METHODS: The Synergy Pocket Micro-pump device (CircuLite, Inc., Saddle Brook, New Jersey) pumps approximately 3.0 l/min, is implanted (off pump) via a mini-thoracotomy, and is positioned in a right subclavicular subcutaneous pocket (like a pacemaker). The inflow cannula inserts into the left atrium; the outflow graft connects to the right subclavian artery.
RESULTS: A total of 17 patients (14 men), age 53 +/- 9 years with ejection fraction 21 +/- 6%, mean arterial pressure 73 +/- 7 mm Hg, pulmonary capillary wedge pressure 29 +/- 6 mm Hg, and cardiac index 1.9 +/- 0.4 l/min/m(2) received an implant. Duration of support ranged from 6 to 213 (median 81) days. In addition to demonstration of significant acute hemodynamic improvements in the first day of support, 9 patients underwent follow-up right heart catheterization at 10.6 +/- 6 weeks. These patients showed significant increases in arterial pressure (67 +/- 8 mm Hg vs. 80 +/- 9 mm Hg, p = 0.01) and cardiac index (2.0 +/- 0.4 l/min/m(2) vs. 2.8 +/- 0.6 l/min/m(2), p = 0.01) with large reductions in pulmonary capillary wedge pressure (30 +/- 5 mm Hg vs. 18 +/- 5 mm Hg, p = 0.001).
CONCLUSIONS: Partial support appears to interrupt the progressive hemodynamic deterioration typical of late-stage heart failure. If proven safe and durable, this device could be used in a relatively large population of patients with severe heart failure who are not sick enough to justify use of currently available full support VADs. (Safety and Performance Evaluation of CircuLite Synergy; NCT00878527).

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19555845     DOI: 10.1016/j.jacc.2009.04.023

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  14 in total

1.  Bridge to recovery: understanding the disconnect between clinical and biological outcomes.

Authors:  Stavros G Drakos; Abdallah G Kfoury; Josef Stehlik; Craig H Selzman; Bruce B Reid; John V Terrovitis; John N Nanas; Dean Y Li
Journal:  Circulation       Date:  2012-07-10       Impact factor: 29.690

2.  Partial cavopulmonary assist from the inferior vena cava to the pulmonary artery improves hemodynamics in failing Fontan circulation: a theoretical analysis.

Authors:  Shuji Shimizu; Toru Kawada; Dai Une; Masafumi Fukumitsu; Michael James Turner; Atsunori Kamiya; Toshiaki Shishido; Masaru Sugimachi
Journal:  J Physiol Sci       Date:  2015-11-06       Impact factor: 2.781

Review 3.  Contemporary strategies in the diagnosis and management of heart failure.

Authors:  Shannon M Dunlay; Naveen L Pereira; Sudhir S Kushwaha
Journal:  Mayo Clin Proc       Date:  2014-03-29       Impact factor: 7.616

4.  The future of adult cardiac assist devices: novel systems and mechanical circulatory support strategies.

Authors:  Carlo R Bartoli; Robert D Dowling
Journal:  Cardiol Clin       Date:  2011-11       Impact factor: 2.213

Review 5.  Management of Heart Failure with Preserved Ejection Fraction: Current Challenges and Future Directions.

Authors:  Bharathi Upadhya; Dalane W Kitzman
Journal:  Am J Cardiovasc Drugs       Date:  2017-08       Impact factor: 3.571

6.  Continuous-flow pump model study: the effect on pump performance of pump characteristics and cardiovascular conditions.

Authors:  Gianfranco Ferrari; Maciej Kozarski; Libera Fresiello; Arianna Di Molfetta; Krzysztof Zieliński; Krystyna Górczyńska; Krzysztof J Pałko; Marek Darowski
Journal:  J Artif Organs       Date:  2013-03-05       Impact factor: 1.731

Review 7.  Heart failure with preserved ejection fraction in the elderly: scope of the problem.

Authors:  Bharathi Upadhya; George E Taffet; Che Ping Cheng; Dalane W Kitzman
Journal:  J Mol Cell Cardiol       Date:  2015-03-06       Impact factor: 5.000

Review 8.  Left ventricular assist device-induced reverse remodeling: it's not just about myocardial recovery.

Authors:  Karolina K Marinescu; Nir Uriel; Douglas L Mann; Daniel Burkhoff
Journal:  Expert Rev Med Devices       Date:  2016-12-22       Impact factor: 3.166

9.  Impact of mechanical unloading on microvasculature and associated central remodeling features of the failing human heart.

Authors:  Stavros G Drakos; Abdallah G Kfoury; Elizabeth H Hammond; Bruce B Reid; Monica P Revelo; Brad Y Rasmusson; Kevin J Whitehead; Mohamed E Salama; Craig H Selzman; Josef Stehlik; Stephen E Clayson; Michael R Bristow; Dale G Renlund; Dean Y Li
Journal:  J Am Coll Cardiol       Date:  2010-07-27       Impact factor: 24.094

Review 10.  Left ventricular assist device unloading effects on myocardial structure and function: current status of the field and call for action.

Authors:  Stavros G Drakos; Abdallah G Kfoury; Craig H Selzman; Divya Ratan Verma; John N Nanas; Dean Y Li; Josef Stehlik
Journal:  Curr Opin Cardiol       Date:  2011-05       Impact factor: 2.161

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.