Literature DB >> 19551547

The efficacy and safety of percutaneous microwave coagulation by a new microwave delivery system in large hepatocellular carcinomas: four case studies.

Zhaocai Yu1, Wenchao Liu, Li Fan, Jie Shao, Ying Huang, Xiaoming Si.   

Abstract

PURPOSE: To demonstrate the efficacy and safety of percutaneous microwave coagulation treatment (PMCT) by a new microwave delivery system (Forsea Microwave) in large hepatocellular carcinomas (HCC) (> or =5 cm).
MATERIALS AND METHODS: Four patients with 4 HCC lesions measuring > or =6 cm in the greatest dimension underwent PMCT by means of the Forsea Microwave microwave delivery system. Final therapeutic efficacy was evaluated with dynamic computer tomography (CT) scans performed within one month after PMCT. During and after PMCT, patients' complaints and any abnormal physical signs were recorded for safety assay. CT or ultrasound scan (US) performed immediately after the treatment was used to detect acute complications related to the treatment. Repeated dynamic CT scans were performed every three to four months thereafter to detect local disease recurrence and/or other recurrences.
RESULTS: Three of these patients achieved a complete ablation of the cancer nodules (two patients with two treatment sessions and one patient with three treatment sessions). One of these patients obtained a complete ablation of the cancer nodule with two treatment sessions except the lesion of portal vein tumour thrombus (PVTT). No obvious symptomatic complication was observed except abdominal pain during and after the treatment in two of these patients. All the patients remained asymptomatic and no recurrent tumour was observed during their follow-up (1-19 months).
CONCLUSIONS: PMCT by the Forsea Microwave microwave delivery system could offer a satisfactory therapeutic effect and is applicable to the treatment of large HCC.

Entities:  

Mesh:

Year:  2009        PMID: 19551547     DOI: 10.1080/02656730902976815

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  3 in total

Review 1.  Microwave tumor ablation: mechanism of action, clinical results, and devices.

Authors:  Meghan G Lubner; Christopher L Brace; J Louis Hinshaw; Fred T Lee
Journal:  J Vasc Interv Radiol       Date:  2010-08       Impact factor: 3.464

Review 2.  Microwave ablation of hepatic malignancy.

Authors:  Meghan G Lubner; Christopher L Brace; Tim J Ziemlewicz; J Louis Hinshaw; Fred T Lee
Journal:  Semin Intervent Radiol       Date:  2013-03       Impact factor: 1.513

3.  The effect of lipiodol deposition in HCC after TACE on the necrosis range of PMCT.

Authors:  HongLiang Sun; JiaYan Ni; XiongYing Jiang; Dong Chen; YaoTing Chen; LinFeng Xu
Journal:  Onco Targets Ther       Date:  2017-08-01       Impact factor: 4.147