| Literature DB >> 19542314 |
Benoît Tyl1, Meriam Kabbaj, Basmah Fassi, Patrick De Jode, William Wheeler.
Abstract
This study compares the ability of 2 semiautomated methods with a fully automated method for QT measurement to minimize the sample size required to detect a moxifloxacin effect and exclude a placebo effect in a thorough QT/QTc study. The fully automated and 1 of 2 semiautomated methods used a global QT measurement in 12 leads, whereas the other semiautomated method used a tangent method on single lead raw complexes. Mean QTcF intervals were greater when measured on a global QT electrocardiogram than on raw complexes, but the mean magnitudes of DeltaQTcF were similar for all methods. The 3 methods detected a statistically significant increase in QTcF for moxifloxacin compared to placebo and were able to exclude a placebo effect on QTcF in all 62 participants. However, due to a smaller variability, the semiautomated methods allowed these detections with fewer than 20 participants, whereas the fully automated required at least 27 participants.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19542314 DOI: 10.1177/0091270009337944
Source DB: PubMed Journal: J Clin Pharmacol ISSN: 0091-2700 Impact factor: 3.126