Computer-aided pulmonary nodule detection - performance of two CAD systems at different CT dose levels.

P A Hein1, P Rogalla, C Klessen, A Lembcke, V C Romano.   

Abstract

PURPOSE: To evaluate the impact of dose reduction on the performance of computer-aided lung nodule detection systems (CAD) of two manufacturers by comparing respective CAD results on ultra-low-dose computed tomography (ULD-CT) and standard dose CT (SD-CT).
MATERIALS AND METHODS: Multi-slice computed tomography (MSCT) data sets of 26 patients (13 male and 13 female, patients 31 - 74 years old) were retrospectively selected for CAD analysis. Indication for CT examination was staging of a known primary malignancy or suspected pulmonary malignancy. CT images were consecutively acquired at 5 mAs (ULD-CT) and 75 mAs (SD-CT) with 120 kV tube voltage (1 mm slice thickness). The standard of reference was determined by three experienced readers in consensus. CAD reading algorithms (pre-commercial CAD system, Philips, Netherlands: CAD-1; LungCARE, Siemens, Germany: CAD-2) were applied to the CT data sets.
RESULTS: Consensus reading identified 253 nodules on SD-CT and ULD-CT. Nodules ranged in diameter between 2 and 41 mm (mean diameter 4.8 mm). Detection rates were recorded with 72 % and 62 % (CAD-1 vs. CAD-2) for SD-CT and with 73 % and 56 % for ULD-CT. Median false positive rates per patient were calculated with 6 and 5 (CAD-1 vs. CAD-2) for SD-CT and with 8 and 3 for ULD-CT. After separate statistical analysis of nodules with diameters of 5 mm and greater, the detection rates increased to 83 % and 61 % for SD-CT and to 89 % and 67 % for ULD-CT (CAD-1 vs. CAD-2). For both CAD systems there were no significant differences between the detection rates for standard and ultra-low-dose data sets (p > 0.05).
CONCLUSION: Dose reduction of the underlying CT scan did not significantly influence nodule detection performance of the tested CAD systems.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19536726     DOI: 10.1055/s-0028-1109394

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  4 in total

1.  Impact of a computer-aided detection (CAD) system integrated into a picture archiving and communication system (PACS) on reader sensitivity and efficiency for the detection of lung nodules in thoracic CT exams.

Authors:  Luca Bogoni; Jane P Ko; Jeffrey Alpert; Vikram Anand; John Fantauzzi; Charles H Florin; Chi Wan Koo; Derek Mason; William Rom; Maria Shiau; Marcos Salganicoff; David P Naidich
Journal:  J Digit Imaging       Date:  2012-12       Impact factor: 4.056

Review 2.  Pulmonary quantitative CT imaging in focal and diffuse disease: current research and clinical applications.

Authors:  Mario Silva; Gianluca Milanese; Valeria Seletti; Alarico Ariani; Nicola Sverzellati
Journal:  Br J Radiol       Date:  2018-01-12       Impact factor: 3.039

Review 3.  A computer-aided diagnosis for evaluating lung nodules on chest CT: the current status and perspective.

Authors:  Jin Mo Goo
Journal:  Korean J Radiol       Date:  2011-03-03       Impact factor: 3.500

4.  Optimal dose levels in screening chest CT for unimpaired detection and volumetry of lung nodules, with and without computer assisted detection at minimal patient radiation.

Authors:  Andreas Christe; Zsolt Szucs-Farkas; Adrian Huber; Philipp Steiger; Lars Leidolt; Justus E Roos; Johannes Heverhagen; Lukas Ebner
Journal:  PLoS One       Date:  2013-12-26       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.