| Literature DB >> 19533622 |
Dirk J Faber1, Ton G van Leeuwen.
Abstract
We present a calibration method for SD-OCT domain spectrometers based on the M-scan of a moving mirror. This method allows determination of the wavenumber sampling increment which determines the depth axis assigned to the structural image. It also allows wavelength calibration of individual pixels which ensures correct re-sampling prior to Fast Fourier Transform. Determination of the spectrometer resolution, which determines sensitivity and resolution decay with depth, is also possible. The wavelength calibration results suggest that hardware calibration of the spectrometers is a necessary complement to the computational methods presented here.Mesh:
Year: 2009 PMID: 19533622 DOI: 10.1002/jbio.200910018
Source DB: PubMed Journal: J Biophotonics ISSN: 1864-063X Impact factor: 3.207