It is important that analytical results, produced to demonstrate compliance with exposure limits are comparable, to ensure controls are monitored to similar standards. Correcting a measurement result of respirable alpha-quartz for the percentage of crystalline material in the calibration dust is good analytical practice and significant changes in the values assigned to calibration materials will affect the interpretation of results by an analyst or occupational hygiene professional. The reissue of the certification for the quartz reference material NIST 1878a in 2005 and differences in comparative values obtained by other work created uncertainty about the values of crystallinity assigned to national calibration dusts for alpha-quartz. Members of an International Organization for Standardization working group for silica measurement ISO/TC146/SC2/WG7 collaborated to investigate the comparability of results by X-ray diffraction (XRD) and to reach a consensus. This paper lists the values recommended by the working group for use with XRD analysis. The values for crystallinity obtained for some of the materials (NIST 1878, Min-U-Sil5 and A9950) were 6-7% lower than the original certification or estimates reported in other comparisons. Crystallinity values obtained by XRD gave a good correlation with BET surface area measurements (r2 = 0.91) but not with mean aerodynamic particle size (r2 = 0.31). Subsamples of two of the materials (A9950 Respirable and Quin 1 Respirable) with smaller particle size distribution than their parent material did not show any significant change in their values for crystallinity, suggesting that the area XRD measurement of these materials within the particle size range collected is more dependent on how the quartz is formed geologically or how it is processed for use. A comparison of results from laboratories using the infrared (IR) and KBr disc method showed that this method is more dependent than XRD on differences in the particle size within the respirable size range, whereas the XRD values were more consistent between the different measurement values obtained on each material. It was not possible to assign a value for percentage purity to each material for users of IR analysis. This work suggests that differences are likely to exist between the results from XRD and IR analysis when measuring 'real' workplace samples and highlights the importance of matching the particle size of the calibration material to the particle size of the workplace dust for measurements of crystalline quartz.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19531809     DOI: 10.1093/annhyg/mep038

Source DB:  PubMed          Journal:  Ann Occup Hyg        ISSN: 0003-4878


  7 in total

1.  A New Miniature Respirable Sampler for In-mask Sampling: Part 1-Particle Size Selection Performance.

Authors:  Peter Stacey; Andrew Thorpe; Rhiannon Mogridge; Taekhee Lee; Martin Harper
Journal:  Ann Occup Hyg       Date:  2016-09-14

2.  Evaluation of Diffuse Reflection Infrared Spectrometry for End-of-Shift Measurement of α-quartz in Coal Dust Samples.

Authors:  Arthur L Miller; Nathaniel C Murphy; Sean J Bayman; Zachary P Briggs; Andrew D Kilpatrick; Courtney A Quinn; Mackenzie R Wadas; Emanuele G Cauda; Peter R Griffiths
Journal:  J Occup Environ Hyg       Date:  2015       Impact factor: 2.155

3.  Evaluation of laser-induced breakdown spectroscopy (LIBS) for measurement of silica on filter samples of coal dust.

Authors:  Christopher B Stipe; Arthur L Miller; Jonathan Brown; Edward Guevara; Emanuele Cauda
Journal:  Appl Spectrosc       Date:  2012-11       Impact factor: 2.388

4.  Deposition Uniformity of Coal Dust on Filters and Its Effect on the Accuracy of FTIR Analyses for Silica.

Authors:  Arthur L Miller; Pamela L Drake; Nathaniel C Murphy; Emanuele G Cauda; Ryan F LeBouf; Gediminas Markevicius
Journal:  Aerosol Sci Technol       Date:  2013-03-07       Impact factor: 2.908

5.  Evaluating portable infrared spectrometers for measuring the silica content of coal dust.

Authors:  Arthur L Miller; Pamela L Drake; Nathaniel C Murphy; James D Noll; Jon C Volkwein
Journal:  J Environ Monit       Date:  2011-12-01

6.  Evaluating the use of a field-based silica monitoring approach with dust from copper mines.

Authors:  Emanuele Cauda; Lauren Chubb; Rustin Reed; Robert Stepp
Journal:  J Occup Environ Hyg       Date:  2018-10       Impact factor: 2.155

7.  An In Vitro Lung System to Assess the Proinflammatory Hazard of Carbon Nanotube Aerosols.

Authors:  Hana Barosova; Bedia Begum Karakocak; Dedy Septiadi; Alke Petri-Fink; Vicki Stone; Barbara Rothen-Rutishauser
Journal:  Int J Mol Sci       Date:  2020-07-27       Impact factor: 5.923

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.