Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle.

L M Sakata1, T T L Wong, H-T Wong, R S Kumar, H M Htoon, H T Aung, M He, T Aung.   

Abstract

AIMS: To compare the diagnostic performance of two anterior segment optical coherence tomography (AS-OCT) devices in assessing the anterior chamber angle (ACA).
METHODS: Visante-OCT and slit-lamp-OCT (SL-OCT) were performed on 101 patients by a single operator. The AS-OCT images were processed by customised 'dewarping' software and assessed by two glaucoma specialists masked to clinical findings. A closed ACA was defined by the presence of contact between the iris and angle anterior to the scleral spur. Measurements of the ACA, anterior chamber depth (ACD), and pupil diameter were analysed. Gonioscopy was performed by another examiner masked to AS-OCT findings.
RESULTS: Qualitative analysis could be carried out in 83 (83%) eyes and quantitative analysis in 61 (60%) eyes. A closed angle in at least one quadrant of the eye was observed in 30 eyes with gonioscopy; Visante-OCT imaging identified 29 of 30 (97%) and SL-OCT imaging identified 27 of 30 (90%) of these eyes (P=0.50, McNemar test). Visante-OCT detected more eyes with at least one closed quadrant than SL-OCT (55 vs46 eyes, respectively, P=0.01). Overall, SL-OCT had better agreement with gonioscopy than with Visante-OCT. Both AS-OCTs showed good agreement for ACD measurements; however, SL-OCT tended to provide consistently higher ACA measurements and smaller pupil diameters than did Visante-OCT.
CONCLUSIONS: Both AS-OCT devices detected most of the eyes with closed ACA on gonioscopy. However, Visante-OCT detected more closed ACAs than did SL-OCT. The better agreement between SL-OCT and gonioscopy is likely because of the use of visible light during both examinations. The ACA measurements obtained with each device are not interchangeable.

Entities:  

Mesh:

Year:  2009        PMID: 19521435     DOI: 10.1038/eye.2009.141

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  13 in total

1.  Comparison of slit lamp-adapted optical coherence tomography features of fellow eyes of acute primary angle closure and eyes with open angle glaucoma.

Authors:  Argyrios Tzamalis; Duy-Thoai Pham; Christopher Wirbelauer
Journal:  Jpn J Ophthalmol       Date:  2014-02-06       Impact factor: 2.447

2.  Comparison of Scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles.

Authors:  D S Grewal; G S Brar; R Jain; S P S Grewal
Journal:  Eye (Lond)       Date:  2011-02-18       Impact factor: 3.775

3.  Anterior segment optical coherence tomography of acute primary angle closure.

Authors:  Hai Tao Zhang; Liang Xu; Wei Fang Cao; Ya Xing Wang; Jost B Jonas
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-02-02       Impact factor: 3.117

4.  Distribution of anterior chamber angle measurements in South African young adults: an optical coherence tomography study.

Authors:  Nishanee Rampersad; Rekha Hansraj
Journal:  Int Ophthalmol       Date:  2022-01-05       Impact factor: 2.029

5.  Non-contact tests for identifying people at risk of primary angle closure glaucoma.

Authors:  Anish Jindal; Irene Ctori; Gianni Virgili; Ersilia Lucenteforte; John G Lawrenson
Journal:  Cochrane Database Syst Rev       Date:  2020-05-28

6.  Comparison of deep anterior lamellar keratoplasty and penetrating keratoplasty with respect to postoperative corneal sensitivity and tear film function.

Authors:  Xiaolei Lin; Binbin Xu; Yifang Sun; Jing Zhong; Weilan Huang; Jin Yuan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-07-31       Impact factor: 3.117

7.  Predictors of Intraocular Pressure After Phacoemulsification in Primary Open-Angle Glaucoma Eyes with Wide Versus Narrower Angles (An American Ophthalmological Society Thesis).

Authors:  Shan C Lin; Marisse Masis; Travis C Porco; Louis R Pasquale
Journal:  Trans Am Ophthalmol Soc       Date:  2017-10-23

Review 8.  Angle imaging: advances and challenges.

Authors:  Desmond T L Quek; Monisha E Nongpiur; Shamira A Perera; Tin Aung
Journal:  Indian J Ophthalmol       Date:  2011-01       Impact factor: 1.848

9.  The role of anterior segment optical coherence tomography in glaucoma.

Authors:  Sarwat Salim
Journal:  J Ophthalmol       Date:  2012-08-01       Impact factor: 1.909

10.  Morphological analysis of age-related iridocorneal angle changes in normal and glaucomatous cases using anterior segment optical coherence tomography.

Authors:  Yuko Maruyama; Kazuhiko Mori; Yoko Ikeda; Morio Ueno; Shigeru Kinoshita
Journal:  Clin Ophthalmol       Date:  2013-12-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.