Delayed cardiac perforation by one small body diameter defibrillator lead.

Majid Haghjoo1, Abolfath Alizadeh, Amir Farjam Fazelifar, Mozhgan Hajahmadi, Mohammad Ali Sadr-Ameli.   

Abstract

We report a 37-year-old man who presented with continuous chest pain 6 weeks after implantable cardioverter-defibrillator implantation. Implantable cardioverter-defibrillator interrogation indicated complete loss of capture even with maximum output. Chest radiography and echocardiography confirmed extracardiac location of lead tip. After lead repositioning in electrophysiology laboratory, acceptable pacing threshold was obtained with no complication. This report demonstrates a case of delayed cardiac perforation after implantation of the St Jude Medical Durata implantable cardioverter-defibrillator lead.

Entities:  

Mesh:

Year:  2010        PMID: 19520384     DOI: 10.1016/j.jelectrocard.2009.04.005

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438