Literature DB >> 19515091

Variability of anti-PF4/heparin antibody results obtained by the rapid testing system ID-H/PF4-PaGIA.

S Schneiter1, G Colucci, I Sulzer, G Barizzi, B Lämmle, L Alberio.   

Abstract

BACKGROUND: Recent studies have shown that a low clinical pretest probability may be adequate for excluding heparin-induced thrombocytopenia. However, for patients with intermediate or high pretest probability, laboratory testing is essential for confirming or refuting the diagnosis. Rapid assessment of anti-PF4/heparin-antibodies may assist clinical decision-making.
OBJECTIVES: To evaluate the performance of rapid ID-H/PF4-PaGIA. In particular, we verified reproducibility of results between plasma and serum specimens, between fresh and frozen samples, and between different ID-H/PF4-polymer lots (polystyrene beads coated with heparin/PF4-complexes). PATIENTS/
METHODS: The samples studied were 1376 plasma and 914 corresponding serum samples from patients investigated for suspected heparin-induced thrombocytopenia between January 2000 and October 2008. Anti-PF4/heparin-antibodies were assessed by ID-H/PF4-PaGIA, commercially available ELISAs and heparin-induced platelet aggregation test.
RESULTS: Among 914 paired plasma/serum samples we noted discordant results (negative vs. low-titre positive) in nine instances (1%; 95%CI, 0.4-1.6%). Overall, agreement between titres assessed in plasma vs. serum was highly significant (Spearman correlation coefficient, 0.975; P < 0.0001). Forty-seven samples tested both fresh and after freezing/thawing showed a good agreement, with one discordant positive/negative result (Spearman correlation coefficient, 0.970; P < 0.0001). Among 1376 plasma samples we noted a strikingly variable incidence of false negative results (none - 82%; 95%CI, 66-98%), depending on the employed ID-H/PF4-polymer lot. Faulty lots can be recognized by titrating commercial positive controls and stored samples of HIT-patients.
CONCLUSION: Laboratories performing the assay should implement stringent internal quality controls in order to recognize potentially faulty ID-H/PF4-polymer lots, thus avoiding false negative results.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19515091     DOI: 10.1111/j.1538-7836.2009.03507.x

Source DB:  PubMed          Journal:  J Thromb Haemost        ISSN: 1538-7836            Impact factor:   5.824


  3 in total

1.  Rapid exclusion or confirmation of heparin-induced thrombocytopenia: a single-center experience with 1,291 patients.

Authors:  Vanessa Nellen; Irmela Sulzer; Gabriela Barizzi; Bernhard Lämmle; Lorenzo Alberio
Journal:  Haematologica       Date:  2011-09-20       Impact factor: 9.941

2.  Rapid immunoassays for diagnosis of heparin-induced thrombocytopenia: Comparison of diagnostic accuracy, reproducibility, and costs in clinical practice.

Authors:  Andriyana Bankova; Yvonne Andres; Michael P Horn; Lorenzo Alberio; Michael Nagler
Journal:  PLoS One       Date:  2017-06-08       Impact factor: 3.240

Review 3.  Challenges in Detecting Clinically Relevant Heparin-Induced Thrombocytopenia Antibodies.

Authors:  Theodore E Warkentin
Journal:  Hamostaseologie       Date:  2020-10-22       Impact factor: 1.778

  3 in total