Literature DB >> 19507679

Comparison of one-dimensional and two-dimensional least-squares strain estimators for phased array displacement data.

Richard G P Lopata1, Hendrik H G Hansen, Maartje M Nillesen, Johan M Thijssen, Chris L De Korte.   

Abstract

In this study, the performances of one-dimensional and two-dimensional least-squares strain estimators (LSQSE) are compared. Furthermore, the effects of kernel size are examined using simulated raw frequency data of a widely-adapted hard lesion/soft tissue model. The performances of both methods are assessed in terms of root-mean-squared errors (RMSE), elastographic signal-to-noise ratio (SNRe) and contrast-to-noise ratio (CNRe). RMSE analysis revealed that the 2D LSQSE yields better results for phased array data, especially for larger insonification angles. Using a 2D LSQSE enabled the processing of unfiltered displacement data, in particular for the lateral/horizontal strain components. The SNRe and CNRe analysis showed an improvement in precision and almost no loss in contrast using 2D LSQSE. However, the RMSE images for different kernel sizes revealed that the optimal 2D kernel size depends on the region-of-interest and showed that the LSQ kernel size should be limited to avoid loss in resolution.

Mesh:

Year:  2009        PMID: 19507679     DOI: 10.1177/016173460903100101

Source DB:  PubMed          Journal:  Ultrason Imaging        ISSN: 0161-7346            Impact factor:   1.578


  6 in total

Review 1.  New developments in paediatric cardiac functional ultrasound imaging.

Authors:  Chris L de Korte; Maartje M Nillesen; Anne E C M Saris; Richard G P Lopata; Johan M Thijssen; Livia Kapusta
Journal:  J Med Ultrason (2001)       Date:  2013-12-20       Impact factor: 1.314

2.  Mapping 3D Strains with Ultrasound Speckle Tracking: Method Validation and Initial Results in Porcine Scleral Inflation.

Authors:  Benjamin Cruz Perez; Elias Pavlatos; Hugh J Morris; Hong Chen; Xueliang Pan; Richard T Hart; Jun Liu
Journal:  Ann Biomed Eng       Date:  2015-11-12       Impact factor: 3.934

3.  Feasibility and relevance of compound strain imaging in non-stenotic arteries: comparison between individuals with cardiovascular diseases and healthy controls.

Authors:  Martijn F H Maessen; Thijs M H Eijsvogels; Ayla Grotens; Maria T E Hopman; Dick H J Thijssen; Hendrik H G Hansen
Journal:  Cardiovasc Ultrasound       Date:  2017-05-18       Impact factor: 2.062

4.  Strain imaging of the lateral collateral ligament using high frequency and conventional ultrasound imaging: An ex-vivo comparison.

Authors:  Kaj Gijsbertse; André Sprengers; Hamid Naghibi Beidokhti; Maartje Nillesen; Chris de Korte; Nico Verdonschot
Journal:  J Biomech       Date:  2018-03-29       Impact factor: 2.712

5.  Long term cerebral and vascular complications after irradiation of the neck in head and neck cancer patients: a prospective cohort study: study rationale and protocol.

Authors:  Joyce Wilbers; Arnoud C Kappelle; Roy Pc Kessels; Stefan Ca Steens; Frederick Ja Meijer; Johannes H Kaanders; Roy Am Haast; Laura E Versteeg; Anil M Tuladhar; Chris L de Korte; Hendrik Hg Hansen; Frank J Hoebers; Willem Boogerd; Erik D van Werkhoven; Marlies E Nowee; Guus Hart; Harry Bartelink; Lucille D Dorresteijn; Ewoud J van Dijk
Journal:  BMC Neurol       Date:  2014-06-19       Impact factor: 2.474

6.  Ultrasound-based estimation of remaining cardiac function in LVAD-supported ex vivo hearts.

Authors:  Louis S Fixsen; Niels J Petterson; Patrick Houthuizen; Marcel C M Rutten; Frans N van de Vosse; Richard G P Lopata
Journal:  Artif Organs       Date:  2020-04-18       Impact factor: 3.094

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.