Investigation of the thermal tissue effects of the argon plasma coagulation modes "pulsed" and "precise" on the porcine esophagus, ex vivo and in vivo.

Georg F Kähler1, Mara N I Szyrach, Ansgar Hieronymus, Rainer Grobholz, Markus D Enderle.   

Abstract

BACKGROUND: Argon plasma coagulation (APC) is a monopolar, noncontact, thermal procedure that is widely used in therapeutic endoscopy. Systematic investigations of the tissue damage ex vivo and in vivo with the new, second-generation APC modes are lacking.
OBJECTIVE: The aim of this study is to compare the tissue effects of the pulsed effect 2 and precise APC modes. DESIGN AND
SETTING: Ex vivo and in vivo animal model.
SUBJECTS: This study involved 3 explanted porcine esophagi and 8 pigs under general anesthesia. INTERVENTION: APC application on 3 explanted esophagi and during esophagoscopy. MAIN OUTCOME MEASUREMENTS: The tissue effect was subjected to histological and statistical investigation.
RESULTS: In vivo, a well known type of superficial tissue damage (type A) of the tunica mucosa and a new injury pattern (type B) limited to the tunica muscularis, were found. Ex vivo, only type A injuries were seen. Thermal injury of the tunica muscularis was significantly lower with precise APC compared with pulsed APC in vivo. The pulsed effect 2 shows a positive correlation between the penetration depth and the power (r = 0.38, P < .0002) or application time for the highest power setting used (40 W, r = 0.77, P < .0001). This correlation could not be detected with precise APC because of its very superficial tissue effect. LIMITATIONS: This was an animal study. The distance of the APC probe to the esophagus may have varied between applications in vivo.
CONCLUSION: Thermal damage by APC of the esophageal tunica muscularis seems to be underestimated ex vivo. The extent of tissue injury was significantly lower with precise APC than with pulsed APC, indicating that precise APC may be suitable for the treatment of particularly thermosensitive, thin-wall anatomy.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19500786     DOI: 10.1016/j.gie.2008.11.050

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  3 in total

1.  Argon and helium plasma coagulation of porcine liver tissue.

Authors:  Tim R Glowka; Jens Standop; Pascal Paschenda; Michael Czaplik; Jörg C Kalff; René H Tolba
Journal:  J Int Med Res       Date:  2017-06-29       Impact factor: 1.671

2.  Esophageal squamous papillomas with focal dermal hypoplasia and eosinophilic esophagitis.

Authors:  Eric A Pasman; Theresa A Heifert; Cade M Nylund
Journal:  World J Gastroenterol       Date:  2017-03-28       Impact factor: 5.742

3.  Trans-Mucosal Efficacy of Non-Thermal Plasma Treatment on Cervical Cancer Tissue and Human Cervix Uteri by a Next Generation Electrosurgical Argon Plasma Device.

Authors:  Thomas Wenzel; Daniel A Carvajal Berrio; Christl Reisenauer; Shannon Layland; André Koch; Diethelm Wallwiener; Sara Y Brucker; Katja Schenke-Layland; Eva-Maria Brauchle; Martin Weiss
Journal:  Cancers (Basel)       Date:  2020-01-22       Impact factor: 6.639

  3 in total