Comparison of a microporous thermoplastic polyurethane-covered stent with a self-expanding bare nitinol stent in a porcine iliac artery model.

Boris Radeleff1, Lars Grenacher, Patricia Christoph, Christof M Sommer, Ulrike Stampfl, Stefanie Ramsauer, Tina Henn, Patrick Kurz, Ruben Lopez-Benitez, Irina Berger, Goetz M Richter.   

Abstract

PURPOSE: To compare outcomes with a thermoplastic polyurethane (TPU)-covered self-expanding nitinol stent-graft (TPU graft) with those of a bare self-expanding nitinol stent in a porcine model.
MATERIALS AND METHODS: Fourteen TPU grafts and 14 commercially available bare nitinol stents were implanted, one each, in the iliac arteries of 14 minipigs. Follow-up was performed at 1 week (six animals), 4 weeks (four animals), and 12 weeks (four animals). The primary study endpoint was in-stent stenosis assessed with quantitative angiography and microscopy. Secondary endpoints were injury, inflammation, and endothelialization.
RESULTS: After 1 week, the maximum percentage luminal loss was significantly greater in TPU grafts (average, 16.2%; range, 0.0%-35.8%) than in bare nitinol stents (8.2%; 0.0%-17.3%) (P = .04). Three of the four TPU grafts were occluded after 4 weeks, and all four TPU grafts were occluded after 12 weeks. Binary stenosis was seen in three of four bare nitinol stents after both 4 and 12 weeks. At 4-week follow-up, the average percentage luminal loss was significantly greater in TPU grafts (85.2%; 40.8%-100%) than in bare nitinol stents (49.5%; 37.9%-62.4%) (P = .003). The difference in neointimal height and percentage average stenosis between TPU grafts (1,028.7 microm and 68.4%) and bare nitinol stents (1,033.6 microm [918.0-1,118.40 microm] and 68.1% [60.44%-71.99%]) was not statistically significant. After 12 weeks, the average percentage luminal loss was 100% in TPU grafts due to occlusion of all stent-grafts and 24.9% (8.0%-63.9%) in bare nitinol stents (P = .011).
CONCLUSIONS: TPU grafts failed to provide improved patency compared with the bare nitinol stents because of excessive neointimal growth and subsequent occlusion. In addition, the bare nitinol stents showed considerable in-stent stenosis at angiography and microscopy.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19497761     DOI: 10.1016/j.jvir.2009.04.044

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Comparison of diamond-like carbon-coated nitinol stents with or without polyethylene glycol grafting and uncoated nitinol stents in a canine iliac artery model.

Authors:  J H Kim; J H Shin; D H Shin; M-W Moon; K Park; T-H Kim; K M Shin; Y H Won; D K Han; K-R Lee
Journal:  Br J Radiol       Date:  2011-03       Impact factor: 3.039

2.  Efficacy of thermoplastic polyurethane and gelatin blended nanofibers covered stent graft in the porcine iliac artery.

Authors:  Dae Sung Ryu; Dong-Sung Won; Ji Won Kim; Yubeen Park; Song Hee Kim; Jeon Min Kang; Chu Hui Zeng; Dohyung Lim; Hyun Choi; Jung-Hoon Park
Journal:  Sci Rep       Date:  2022-10-03       Impact factor: 4.996

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.