Clinical impact of leak compensation during non-invasive ventilation.

Jan Hendrik Storre1, Philipp Bohm, Michael Dreher, Wolfram Windisch.   

Abstract

BACKGROUND: This study aimed to assess the impact of leak compensation capabilities during pressure- and volume-limited non-invasive positive-pressure ventilation (NPPV) in COPD patients.
METHODS: Fourteen patients with stable hypercapnic COPD who were receiving long-term NPPV were included in the study. For both modes of NPPV, a full face mask and an artificial leak in the ventilatory circuit were used at three different settings, and applied during daytime NPPV, either without leakage (setting I), with leakage during inspiration only (setting II), and with leakage during inspiration and expiration (setting III). Ventilation pattern was pneumotachy-graphically recorded.
RESULTS: NPPV was feasible with negligible leak volumes, indicating optimal mask fitting during the daytime (setting I). In the presence of leakage (settings II and III), the attempt to compensate for leak was only evident during pressure-limited NPPV, since inspiratory volumes delivered by the ventilator increased from 726+/-129 (setting I) to 1104+/-164 (setting II), and to 1257+/-166 (setting III) ml during pressure-limited NPPV, respectively (all p<0.001); however, they remained stable during volume-limited NPPV. Leak compensation resulted in a decrease in leakage-induced dyspnea. However, 83%/87% (setting II/III) of the additionally-delivered inspiratory volume during pressure-limited NPPV was also lost via leakage. Expiratory volume was higher in setting II compared to setting III (both p<0.001), indicating the presence of significant expiratory leakage.
CONCLUSIONS: The attempt at leak compensation largely feeds the leakage itself and only results in a marginal increase of tidal volume. However, pressure-limited--but not volume-limited--NPPV results in a clinically-important leak compensation in vivo. TRIAL REGISTRATION: www.uniklinik-freiburg.de/zks/live/uklregister/Oeffentlich.html Identifier: UKF001272.

Entities:  

Mesh:

Year:  2009        PMID: 19481433     DOI: 10.1016/j.rmed.2009.04.020

Source DB:  PubMed          Journal:  Respir Med        ISSN: 0954-6111            Impact factor:   3.415


  3 in total

Review 1.  Basic Concepts for Tidal Volume and Leakage Estimation in Non-Invasive Ventilation.

Authors:  Manel Luján; Cristina Lalmolda; Begüm Ergan
Journal:  Turk Thorac J       Date:  2019-04-01

2.  Indirect calorimetry during non invasive mechanical ventilation. Is the next step for gas exchange monitoring?

Authors:  Antonio M Esquinas; Güniz Koksal
Journal:  J Clin Monit Comput       Date:  2012-10-16       Impact factor: 2.502

Review 3.  Comfort During Non-invasive Ventilation.

Authors:  Gianmaria Cammarota; Rachele Simonte; Edoardo De Robertis
Journal:  Front Med (Lausanne)       Date:  2022-03-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.