| Literature DB >> 19472627 |
Jagdish P Bhatnagar1, Josef Novotny, Mubina A Quader, Greg Bednarz, M Saiful Huq.
Abstract
The calibration of Leksell Gamma Knife Perfexion (LGK PFX) is performed using a spherical polystyrene phantom 160 mm in diameter, which is provided by the manufacturer. This is the same phantom that has been used with LGK models U, B, C, and 4C. The polystyrene phantom is held in irradiation position by an aluminum adaptor, which has stainless steel side-fixation screws. The phantom adaptor partially attenuates the beams from sectors 3 and 7 by 3.2% and 4.6%, respectively. This unintended attenuation introduces a systematic error in dose calibration. The overall effect of phantom-adaptor attenuation on output calibration of the LGK PFX unit is to underestimate output by about 1.0%.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19472627 DOI: 10.1118/1.3093240
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071