Glutest Neo Super--a new handheld blood glucose meter-corrects for the effects of the hematocrit values in both hematocrit-adjusted samples and samples obtained from anemic patients.

Fumiatsu Yakushiji1, Hiroshi Fujita, Hiroshi Suzuki, Ritsuko Joukyu, Mutsuko Yasuda, Yoshiyasu Terayama, Kaoru Nagasawa, Akira Ohwada, Ken Taniguchi, Kazuhiko Fujiki, Masako Shimojo, Hiroyuki Kinoshita.   

Abstract

BACKGROUND: Handheld blood glucose (BG) meters are convenient tools that are widely used to measure the BG levels. However, the hematocrit (Hct) value has been identified as a confounding factor for accurate BG measurement. Some BG meters are equipped with an Hct-correcting feature, whose effectiveness has been tested previously. Nevertheless, the measurements yielded by many BG meters are confounded by the Hct values. Recently, a new BG meter equipped with an Hct-correcting feature has been developed; however, its effectiveness has not yet been confirmed. STUDY
DESIGN: Venous blood samples were collected from two healthy volunteers, and the Hct values in the samples were adjusted to approximately 0%, 10%, 20%, 30%, 40%, 50%, and 60%. Further, venous blood samples were collected from 10 anemic patients (Hct <40%). The whole BG (WBG) levels in the samples were measured using two devices-the new BG meter (Glutest Neo Super [Sanwa Kagaku Kenkyusho Co. Ltd., Nagoya, Japan]) and a standard BG meter (OneTouch Ultra [Life Scan Inc., Milpitas, CA]). For reference, plasma glucose (PG) levels were measured using a machine at our hospital laboratory (GA08 [A&T Co., Kanagawa, Japan]). The bias in the measurements was calculated as follows: bias = ([WBG - PG]/PG) x 100. Further, the correlation between the Hct values and the bias was assessed by performing linear regression analysis.
RESULTS: In both the Hct-adjusted samples and the samples obtained from anemic patients, the WBG levels measured using Glutest Neo Super were minimally affected by the Hct values, while those measured using OneTouch Ultra were affected by the Hct values to a statistically significant extent.
CONCLUSIONS: The Hct-correcting feature of the new BG meter Glutest Neo Super was effective. The use of this new device for BG measurements may lead to more appropriate treatment selection.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19459765     DOI: 10.1089/dia.2008.0095

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  2 in total

1.  Life Under Hypoxia Lowers Blood Glucose Independently of Effects on Appetite and Body Weight in Mice.

Authors:  Sameer Abu Eid; Martina T Hackl; Mairam Kaplanian; Max-Paul Winter; Doris Kaltenecker; Richard Moriggl; Anton Luger; Thomas Scherer; Clemens Fürnsinn
Journal:  Front Endocrinol (Lausanne)       Date:  2018-08-28       Impact factor: 5.555

2.  Reagentless Sensing of Vancomycin Using an Indium Tin Oxide Electrode Grafted with Molecularly Imprinted Polymer including Ferrocenyl Group.

Authors:  Haruto Eguchi; Akihiko Hatano; Yasuo Yoshimi
Journal:  Sensors (Basel)       Date:  2021-12-13       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.