Evaluation of analytical performance of the Pathfast cardiac troponin I.

Francesca Di Serio1, Marco Caputo, Martina Zaninotto, Cosimo Ottomano, Mario Plebani.   

Abstract

BACKGROUND: Cardiac troponins are considered the cornerstone for risk stratification and diagnosis of patients with acute coronary syndrome (ACS). Following Clinical Laboratory Standards Institute (CLSI) guidelines, we assessed the analytical performances of the Pathfast (Mitsubishi, Japan) cTnI method.
METHODS: We evaluated different sample types. Control materials and lithium heparin plasma pools were used to determine: limit of blank (LoB), limit of detection (LoD), imprecision and linearity. The effects of potential endogenous interfering substances and the possibility of falsely increased cardiac troponin I (cTnI) concentrations attributable to the presence of heterophilic antibodies (HA), rheumatoid factor (RF) and human anti-mouse antibodies (HAMA) in high concentrations were evaluated. The 99th percentile limit of the cTnI value distribution was determined from 320 Caucasian reference individuals.
RESULTS: No significant differences were found when cTnI concentrations of 40 lithium-heparin plasma samples were compared with the matched values of K(2)-EDTA plasma, whole blood and serum samples. The LoB and the LoD of the cTnI method were 0.0048 and 0.0066 microg/L, respectively. cTnI mean values from 0.66 to 6.0 microg/L showed a total CV% from 6.0 to 6.4. cTnI at a concentration of 0.02 microg/L was associated with a total CV of 9.6%. The method gave a linear response for cTnI concentrations within the measurement range. In six of 12 samples containing HA, a positive interference was demonstrated. The 99th percentile limit of the cTnI distribution in the reference population was 0.013 microg/L.
CONCLUSIONS: The data indicate that the cTnI Pathfast method may be suitable for helping clinicians in the management of patients with ACS.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19453290     DOI: 10.1515/CCLM.2009.182

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

Review 1.  Effectiveness of practices for improving the diagnostic accuracy of Non ST Elevation Myocardial Infarction in the Emergency Department: A Laboratory Medicine Best Practices™ systematic review.

Authors:  Christopher Layfield; John Rose; Aaron Alford; Susan R Snyder; Fred S Apple; Farah M Chowdhury; Michael C Kontos; L Kristin Newby; Alan B Storrow; Milenko Tanasijevic; Elizabeth Leibach; Edward B Liebow; Robert H Christenson
Journal:  Clin Biochem       Date:  2015-02-07       Impact factor: 3.281

Review 2.  High-sensitivity assays for troponin in patients with cardiac disease.

Authors:  Dirk Westermann; Johannes Tobias Neumann; Nils Arne Sörensen; Stefan Blankenberg
Journal:  Nat Rev Cardiol       Date:  2017-04-06       Impact factor: 32.419

3.  Performance of ivisen IA-1400, a new point-of-care device with an internal centrifuge system, for the measurement of cardiac troponin I levels.

Authors:  Ha Nui Kim; Soo-Young Yoon
Journal:  J Clin Lab Anal       Date:  2021-03-17       Impact factor: 2.352

Review 4.  The Emerging Role of Presepsin (P-SEP) in the Diagnosis of Sepsis in the Critically Ill Infant: A Literature Review.

Authors:  Chiara Maddaloni; Domenico Umberto De Rose; Alessandra Santisi; Ludovica Martini; Stefano Caoci; Iliana Bersani; Maria Paola Ronchetti; Cinzia Auriti
Journal:  Int J Mol Sci       Date:  2021-11-10       Impact factor: 5.923

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.