Same-patient reproducibility of state entropy: a comparison of simultaneous bilateral measurements during general anesthesia.

Mehmet S Ozcan1, David M Thompson, Jorge Cure, J Randal Hine, Pamela R Roberts.   

Abstract

BACKGROUND: State Entropy (SE) is an index of anesthetic depth similar to Bispectral Index (BIS). Both indices use a single-channel electroencephalogram, recorded from a unilaterally applied electrode on the forehead, as their input. Intrapatient reproducibility of BIS was questioned in a recent study in which simultaneous measurements from two electrodes applied to the same patient showed conflicting anesthetic depths. Our purpose was to determine whether SE results are similarly reproducible, even though their computation uses a different algorithm than BIS. In this study, we investigated the reproducibility of SE measurements simultaneously obtained from bilaterally applied electrodes in the same patient.
METHODS: Entropy electrodes were applied bilaterally on 21 patients under general inhaled anesthesia. Simultaneous SE measurements from both electrodes were recorded every 10 s from each patient. Data were analyzed with Bland-Altman statistics.
RESULTS: We obtained 14,379 pairs of SE measurements. Four percent of the individual measurements suggested conflicting anesthetic depth along with a numeric difference more than 10 points. Bias was not clinically significant (-0.3). Ninety-five percent limits of agreement were -11.7 and +11.6.
CONCLUSIONS: SE showed a clinically significant degree of disagreement when probes were applied on both sides of the forehead in the same patient. Bland-Altman statistics showed better same-patient reproducibility in SE than did a similar study on BIS. Nevertheless, 4% of the simultaneously measured pairs of SE suggested different anesthetic depths and differed by more than 10 points. Caution is advised when using SE as a clinical index of anesthetic depth.

Entities:  

Mesh:

Year:  2009        PMID: 19448208     DOI: 10.1213/ane.0b013e31819dc668

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  3 in total

1.  Determining entropy values equivalent to the bispectral index values during sevoflurane anaesthesia.

Authors:  Canan Balcı; Hamit Karabekir; Remziye Sivaci
Journal:  Arch Med Sci       Date:  2010-06-30       Impact factor: 3.318

2.  PRISMA: accuracy of response entropy and bispectral index to predict the transition of consciousness during sevoflurane anesthesia: A systematic review and meta-analysis.

Authors:  Tao Liang; Fan Wu; Baoguo Wang; Feng Mu
Journal:  Medicine (Baltimore)       Date:  2021-04-30       Impact factor: 1.817

3.  Variations in Values of State, Response Entropy and Haemodynamic Parameters Associated with Development of Different Epileptiform Patterns during Volatile Induction of General Anaesthesia with Two Different Anaesthetic Regimens Using Sevoflurane in Comparison with Intravenous Induct: A Comparative Study.

Authors:  Michał Stasiowski; Anna Duława; Izabela Szumera; Radosław Marciniak; Ewa Niewiadomska; Wojciech Kaspera; Lech Krawczyk; Piotr Ładziński; Beniamin Oskar Grabarek; Przemysław Jałowiecki
Journal:  Brain Sci       Date:  2020-06-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.