Literature DB >> 19439482

MR angiographic follow-up of intracranial aneurysms treated with detachable coils: evaluation of a blood-pool contrast medium.

T Kau1, J Gasser, S Celedin, E Rabitsch, W Eicher, E Uhl, K A Hausegger.   

Abstract

BACKGROUND AND
PURPOSE: Blood-pool agents are promising in the imaging of small vessels with slow or complex flow. Our aim was to compare blood-pool contrast-enhanced MR angiography (BPCE-MRA) using gadofosveset trisodium (Vasovist) with 3D time-of-flight MRA (TOF-MRA) in the follow-up of intracranial aneurysms after endovascular therapy.
MATERIALS AND METHODS: We included 32 patients with a total of 37 coiled aneurysms. MRAs in the early steady-state phase were performed on a 1.5T scanner within 8 days of digital subtraction angiography (DSA). Two radiologists independently analyzed TOF-MRA and BPCE-MRA images. Consensus was reached by review involving a third neuroradiologist. DSA images were interpreted separately by an interventional radiologist. Findings were assigned to 1 of 3 categories: 1) complete occlusion, 2) residual neck, and 3) residual aneurysm.
RESULTS: Follow-up DSA demonstrated 13 complete obliterations (class 1), 13 residual necks (class 2), and 11 residual aneurysms (class 3). Weighted kappa statistics showed substantial concordance of TOF-MRA and DSA (0.664) as well as BPCE-MRA and DSA (0.724) ratings. Comparison between TOF-MRA and BPCE-MRA found excellent agreement (0.818) with only 6 (16.2%) discrepancies. For detecting residual flow, the difference in accuracy of both MRA techniques (83.8% versus 91.9%) was not significant (McNemar, P = 1.000). BPCE-MRA showed a tendency towards higher sensitivity and specificity (91.7% and 92.3%, respectively) compared with TOF-MRA (87.5% and 76.9%).
CONCLUSIONS: In classifying the completeness of endovascular cerebral aneurysm therapy, we found that BPCE-MRA and 3D TOF-MRA showed very good agreement. The use of Vasovist did not lead to a significantly increased accuracy of MRA follow-up.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19439482     DOI: 10.3174/ajnr.A1622

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  3 in total

Review 1.  MRA versus DSA for follow-up of coiled intracranial aneurysms: a meta-analysis.

Authors:  M J van Amerongen; H D Boogaarts; J de Vries; A L M Verbeek; F J A Meijer; M Prokop; R H M A Bartels
Journal:  AJNR Am J Neuroradiol       Date:  2013-09-05       Impact factor: 3.825

2.  Follow-up of coiled intracranial aneurysms: comparison of 3D time-of-flight and contrast-enhanced magnetic resonance angiography at 3T in a large, prospective series.

Authors:  Laurent Pierot; Christophe Portefaix; Anne Boulin; Jean-Yves Gauvrit
Journal:  Eur Radiol       Date:  2012-05-09       Impact factor: 5.315

3.  Follow-up after embolization of ruptured intracranial aneurysms: a prospective comparison of two-dimensional digital subtraction angiography, three-dimensional digital subtraction angiography, and time-of-flight magnetic resonance angiography.

Authors:  Zbigniew Serafin; Piotr Strześniewski; Władysław Lasek; Wojciech Beuth
Journal:  Neuroradiology       Date:  2012-04-10       Impact factor: 2.804

  3 in total