Literature DB >> 19427563

Noise reduction in low-dose x-ray fluoroscopy for image-guided radiation therapy.

Jing Wang1, Lei Zhu, Lei Xing.   

Abstract

PURPOSE: To improve the quality of low-dose X-ray fluoroscopic images using statistics-based restoration algorithm so that the patient fluoroscopy can be performed with reduced radiation dose. METHOD AND MATERIALS: Noise in the low-dose fluoroscopy was suppressed by temporal and spatial filtering. The temporal correlation among neighboring frames was considered by the Karhunen-Loève (KL) transform (i.e., principal component analysis). After the KL transform, the selected neighboring frames of fluoroscopy were decomposed to uncorrelated and ordered principal components. For each KL component, a penalized weighted least-squares (PWLS) objective function was constructed to restore the ideal image. The penalty was chosen as anisotropic quadratic, and the penalty parameter in each KL component was inversely proportional to its corresponding eigenvalue. Smaller KL eigenvalue is associated with the KL component of lower signal-to-noise ratio (SNR), and a larger penalty parameter should be used for such KL component. The low-dose fluoroscopic images were acquired using a Varian Acuity simulator. A quality assurance phantom and an anthropomorphic chest phantom were used to evaluate the presented algorithm.
RESULTS: In the images restored by the proposed KL domain PWLS algorithm, noise is greatly suppressed, whereas fine structures are well preserved. Average improvement rate of SNR is 75% among selected regions of interest. Comparison studies with traditional techniques, such as the mean and median filters, show that the proposed algorithm is advantageous in terms of structure preservation.
CONCLUSIONS: The proposed noise reduction algorithm can significantly improve the quality of low-dose X-ray fluoroscopic image and allows for dose reduction in X-ray fluoroscopy.

Entities:  

Mesh:

Year:  2009        PMID: 19427563     DOI: 10.1016/j.ijrobp.2009.01.020

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  5 in total

1.  Scatter correction for full-fan volumetric CT using a stationary beam blocker in a single full scan.

Authors:  Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

2.  Noise reduction for curve-linear structures in real time fluoroscopy applications using directional binary masks.

Authors:  Martin Wagner; Pengfei Yang; Sebastian Schafer; Charles Strother; Charles Mistretta
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

3.  A motion-compensated image filter for low-dose fluoroscopy in a real-time tumor-tracking radiotherapy system.

Authors:  Naoki Miyamoto; Masayori Ishikawa; Kenneth Sutherland; Ryusuke Suzuki; Taeko Matsuura; Chie Toramatsu; Seishin Takao; Hideaki Nihongi; Shinichi Shimizu; Kikuo Umegaki; Hiroki Shirato
Journal:  J Radiat Res       Date:  2014-08-16       Impact factor: 2.724

4.  Half-Fan-Based Intensity-Weighted Region-of-Interest Imaging for Low-Dose Cone-Beam CT in Image-Guided Radiation Therapy.

Authors:  Boyeol Yoo; Kihong Son; Rizza Pua; Jinsung Kim; Alexander Solodov; Seungryong Cho
Journal:  Healthc Inform Res       Date:  2016-10-31

5.  Edge-enhancement densenet for X-ray fluoroscopy image denoising in cardiac electrophysiology procedures.

Authors:  Yimin Luo; Yingliang Ma; Hugh O' Brien; Kui Jiang; Vikram Kohli; Sesilia Maidelin; Mahrukh Saeed; Emily Deng; Kuberan Pushparajah; Kawal S Rhode
Journal:  Med Phys       Date:  2022-01-18       Impact factor: 4.506

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.