Comparison between 20G and 22G needle for second trimester amniocentesis in terms of technical aspects and short-term complications.

Apostolos P Athanasiadis1, Konstantinos Pantazis, Dimitrios G Goulis, Konstantinos Chatzigeorgiou, Violeta Vaitsi, Efstratios Assimakopoulos, Filippos Tzevelekis, Tryfon Tsalikis, John N Bontis.   

Abstract

OBJECTIVE: To compare short-term complications of amniocentesis using 20G versus 22G needle.
METHODS: A total of 200 women referred for mid-trimester amniocentesis were randomized to a 20G (Group I, n = 100) or 22G amniocentesis needle (Group II, n = 100). The primary outcome was intrauterine bleeding at needle insertion. The operator reported technical aspects and patient's reactions immediately after the procedure. Women's perception of discomfort 30 min after the procedure and complications after 2 weeks were recorded.
RESULTS: Intrauterine bleeding at needle insertion was similar between groups (4/100 vs 8/100). When only transplacental taps were analyzed, bleeding was significantly lower in Group I (4/20 vs 8/14, p = 0.035). Fluid retrieval was faster in Group I (9.6 vs 26.8 sec, p < 0.001). In all, 65% of women in Group I versus 30% in Group II reported discomfort during the procedure, although discomfort 30 min after the procedure and complications within 2 weeks after the procedure were similar in the two groups.
CONCLUSION: Amniocentesis with 20G needle is associated with lower risk of intrauterine bleeding in case of transplacental needle insertion and allows for faster fluid retrieval, as compared with 22G needle. Nevertheless, 20G needle is associated with more immediate discomfort during the procedure.

Entities:  

Mesh:

Year:  2009        PMID: 19412914     DOI: 10.1002/pd.2283

Source DB:  PubMed          Journal:  Prenat Diagn        ISSN: 0197-3851            Impact factor:   3.050


  3 in total

1.  Comparison of perinatal outcome after 2nd trimester amniocentesis using 22G and 20G needle.

Authors:  A Athanasiadis; P Petrakis; T Mikos
Journal:  Hippokratia       Date:  2017 Jan-Mar       Impact factor: 0.471

2.  Does bleeding affect fetal Doppler parameters during genetic amniocentesis?

Authors:  Cantekin Iskender; Ebru Tarım; Tayfun Cok; Hakan Kalaycı; Ayşe Parlakgümüş; Cem Yalçınkaya
Journal:  J Turk Ger Gynecol Assoc       Date:  2014-06-01

3.  Conventional 22- and 20-gauge needle for second trimester amniocentesis: A comparison of short term outcomes.

Authors:  Seyit Ali Köse; Mehmet Özgür Akkurt; And Yavuz; Serenat Eriş; Esra Nur Tola; Mehmet Okan Özkaya; Mekin Sezik
Journal:  Turk J Obstet Gynecol       Date:  2016-03-10
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.