Literature DB >> 19408984

Overdrainage of cerebrospinal fluid caused by detachment of the pressure control cam in a programmable valve after 3-tesla magnetic resonance imaging.

Atsushi Watanabe1, Tatsuya Seguchi, Kazuhiro Hongo.   

Abstract

The authors report a rare case of overdrainage of the CSF caused by the malfunction of a Codman-Hakim programmable valve (CHPV) following a 3-T MR imaging procedure. Nine years ago this 72-year-old woman underwent ventriculoperitoneal shunt placement with a CHPV system for hydrocephalus due to subarachnoid hemorrhage. The postoperative course was uneventful and the system functioned well. A radiograph obtained immediately after 3-T MR imaging revealed that the pressure control cam in the valve system was detached from the base plate. Intracranial hypotension syndrome occurred several hours after the MR imaging study, and a CT scan revealed a decrease in ventricle size. A revision of the system promptly resolved the symptoms, and a postoperative CT scan revealed that the ventricle size was restored to normal. Examination of the extracted valve showed a Y-shaped crack in the plastic housing as well as detachment of the white marker and cam from the base plate. A reduction in the power of the flat spring to press the valve ball led to CSF overdrainage because of a loss of support by the cam. Because the patient had incurred no head injury during the day and radiographic studies of the system 5 years previously had shown detachment of the white marker, damage to the system might have been caused by a past impact. These facts may indicate that the antimagnetic performance of the system could have decreased due to a previous impact and that the strong magnetic force in a 3-T MR imaging environment might have caused detachment of the cam.

Entities:  

Mesh:

Year:  2010        PMID: 19408984     DOI: 10.3171/2008.3.17688

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  2 in total

1.  Shunt Overdrainage Caused by Displacement of the Pressure Control Cam after Pressure Adjustment.

Authors:  Su-Ho Kim; Yoon-Soo Lee; Min-Seok Lee; Sang-Jun Suh; Jeong-Ho Lee; Dong-Gee Kang
Journal:  Korean J Neurotrauma       Date:  2016-10-31

2.  In vitro testing of explanted shunt valves in hydrocephalic patients with suspected valve malfunction.

Authors:  Christoph Bettag; Christian von der Brelie; Florian Baptist Freimann; Ulrich-Wilhelm Thomale; Veit Rohde; Ingo Fiss
Journal:  Neurosurg Rev       Date:  2021-05-24       Impact factor: 2.800

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.