A clinical evaluation of the Endocater--an electronic apex locator.

M E Keller1, C E Brown, C W Newton.   

Abstract

The purpose of this study was to evaluate the accuracy of an electronic apex locator, the Endocater, in determining the location of the apical constriction or cementodentinal junction (CDJ). Measurements made by the Endocater were also compared with the adjustments recommended by an experienced endodontist. A total of 69 teeth with 99 canals from nine patients were evaluated. The results indicated that the evaluator was 95.8% and the Endocater 67.7% accurate in positioning the probe within +/- 1.000 mm from the CDJ. Thirty-three canals or 34.4% appeared to be clinically acceptable radiographically and no adjustments were deemed necessary by the evaluator. There was a significant difference (p less than 0.003) between the evaluator's ability to adjust the electronically generated file length radiograph to acceptable clinical standards and the Endocater's ability to determine the location of the CDJ. A total of 59 canals (59.6%) of the electronically determined working lengths were beyond the CDJ (n = 99). In 27 canals, the tip of the probe was greater than 1 mm beyond the constriction. Without radiographs to confirm working lengths, many canals in this study would have been overinstrumented. Future research with electronic apex locators is needed before accepting the technique as a substitution for radiographic working length determination.

Entities:  

Mesh:

Year:  1991        PMID: 1940752     DOI: 10.1016/S0099-2399(06)81865-0

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  3 in total

1.  Comparison of third generation versus fourth generation electronic apex locators in detecting apical constriction: An in vivo study.

Authors:  D V Swapna; Akash Krishna; Anand C Patil; K Rashmi; Veena S Pai; M A Ranjini
Journal:  J Conserv Dent       Date:  2015 Jul-Aug

2.  Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex.

Authors:  Shahriar Shahi; Saeed Rahimi; Amin Salem Milani; Mohammad Asghari Jafarabadi; Gholam Reza Djoorabbaf Naghsh
Journal:  Iran Endod J       Date:  2011-05-15

3.  In Vitro Comparison of Raypex 6 and Endopilot Using a Novel, Computer-Aided Measurement System, for Determining the Working Length.

Authors:  David Christofzik; Falk Schwendicke; Christian Flörke; Alexander Härtl; Christof Dörfer; Birte Größner-Schreiber
Journal:  PLoS One       Date:  2015-08-03       Impact factor: 3.240