Pulse oximetry using a disposable finger sensor placed on the forehead in hypoxic patients.

Howard A Smithline1, Ned Rudnitzky, Sandy Macomber, Fidela S J Blank.   

Abstract

BACKGROUND: Disposable pulse oximetry sensors designed for the finger are sometimes placed on the forehead. Although finger sensors have not been designed to accurately measure arterial oxygen saturation when placed on other body parts, they are used on the forehead when an oximetry waveform cannot be obtained from the finger.
OBJECTIVES: The purpose of this prospective observational study was to measure the agreement between disposable transmittal sensors placed on the finger and the same sensors placed on the forehead.
METHODS: We used a convenience sample of hypoxic emergency department patients. Patients were attached to three oximeters: 1) a non-disposable finger sensor (reference); 2) a disposable finger sensor placed on the finger; and 3) a disposable finger sensor placed on the forehead. Data were analyzed using the Bland-Altman method. A difference of 5% was considered clinically significant.
RESULTS: Twenty-five patients were approached, and 20 were enrolled. Eleven of the patients had an absolute difference between the disposable finger sensor placed on the finger compared to the forehead > or = 5%.
CONCLUSION: Pulse oximetry measurements taken on the forehead using a disposable finger sensor were inaccurate in over half of the subjects. Therefore, disposable finger oximetry sensors should not be placed on the forehead. Copyright 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19406607     DOI: 10.1016/j.jemermed.2009.03.019

Source DB:  PubMed          Journal:  J Emerg Med        ISSN: 0736-4679            Impact factor:   1.484


  2 in total

1.  Pulse oximetry and high-dose vasopressors: a comparison between forehead reflectance and finger transmission sensors.

Authors:  Nicolas Nesseler; Jean-Vincent Frénel; Yoann Launey; Jeff Morcet; Yannick Mallédant; Philippe Seguin
Journal:  Intensive Care Med       Date:  2012-08-07       Impact factor: 17.440

2.  A comparison of finger and forehead pulse oximeters in heart failure patients during maximal exercise.

Authors:  Kevin L Kelly; Alex R Carlson; Thomas G Allison; Bruce D Johnson
Journal:  Heart Lung       Date:  2019-11-14       Impact factor: 2.210

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.