| Literature DB >> 19405650 |
T Marrodán Undagoitia1, F von Feilitzsch, L Oberauer, W Potzel, A Ulrich, J Winter, M Wurm.
Abstract
The fluorescence decay-time constants have been measured for several scintillator mixtures based on phenyl-o-xylylethane (PXE) and linear alkylbenzene (LAB) solvents. The resulting values are of relevance for the physics performance of the proposed large-volume liquid scintillator detector Low Energy Neutrino Astronomy (LENA). In particular, the impact of the measured values to the search for proton decay via p-->K(+)nu is evaluated in this work.Entities:
Year: 2009 PMID: 19405650 DOI: 10.1063/1.3112609
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523