Stephen A Brennan1, Raymond J Walls, Elizabeth Smyth, Talal Al Mulla, John M O'Byrne.
Abstract
BACKGROUND ANDEntities:
Mesh:
Year: 2009 PMID: 19404812 PMCID: PMC2823160 DOI: 10.3109/17453670902930016
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Results of culture from swabs from tourniquets and exsanguinators with mean number of colony forming units (CFUs) before and after decontamination
| Swabs positive for bacterial growth | CFUs/swab before | CFUs/swab after | |
|---|---|---|---|
| Hospital A | 14/28 | 5.3 | 0 |
| Hospital B | 17/28 | 15 | 0.04 |
| Hospital C | 18/28 | 21 | 0.5 |
| Hospital A | 10/20 | 2.3 | 0 |
| Hospital B | 12/20 | 8.8 | 0.05 |
| Hospital C | 11/20 | 2.8 | 0.3 |
| Hospital A | 4/8 | 13 | 0 |
| Hospital B | 5/8 | 31 | 0 |
| Hospital C | 7/8 | 65 | 1.1 |
Comparison of organisms isolated from infected lower limb cases treated in the elective orthopedic hospital with the organisms isolated from exsanguinators and tourniquets. Numbers of colonies identified are given in parentheses
| Database | Exsanguinators and tourniquets |
|---|---|
| Coagulase negative staphylococci 39% | Coagulase-negative staphylococci |
| Methicillin-sensitive | |
| Methicillin-resistant | |
| Methicillin-sensitive | |
These organisms were isolated only from exsanguinators.
Manufacturer’s recommendations for maintenance and storage of the Rhys-Davies exsanguinator
| Ensure that the product is clean and suitable before next application |
| Pressure of the exsanguinator should be equivalent to 60 mmHg |
| The exsanguinator may be washed using soap and water |
| It should not come into contact with solvents |
| Visually inspect device for signs of cracking or splitting, and if so refer to supplier |
| The recommended period of use of the device is 12 months |
| It should be stored in its blue polyethylene bag and placed in its cardboard box |
| Storage temperature 15 ± 10°C |
| It should be stored away from electrical equipment |