Comparative study of accuracy and clinical agreement of the CoaguChek XS portable device versus standard laboratory practice in unexperienced patients.

Eduardo G Torreiro1, Elizabeth Gómez Fernández, Rosa Mariño Rodríguez, Carmen Vázquez López, Julia Barreal Núñez.   

Abstract

The objective of the study was to compare the accuracy and clinical agreement of the CoaguChek XS versus the standard laboratory practice. Forty-one patients on long-term anticoagulation with acenocumarol without previous experience in self-monitoring participated to obtain 218 pairs of data. Several methods for comparative statistics were applied to assess the possible disagreements between techniques as well as a range of previously published criteria of clinical agreement and the very recently described error-grid for INR comparison that we partially modify. The mean age was 52.1 and the indications for oral anticoagulation were prosthetic valves (36.59%), atrial fibrillation (34.15%), venous thromboembolic disease (21.95%) and others (7.31%) with a target range of 2-3 INR units (63.4%) or 2.5-3.5 (36.6%). Analyzing the whole series of data, the Pearsons rho correlation coefficient for precision between methods was 0.95 and the C(b) bias correction factor for accuracy 0.99 with a minimal bias of 0.1 INR units between methods applying the Bland-Altman plot. The linear regression procedure described by Passing and Bablok showed a minimal deviation from the best-fit line and a slope of 0.90. The mean of the absolute relative differences was 7% which is in the "very good" range of agreement. No results were found in the clinically "dangerous" D zone of the error-grids with 99% of data in the clinically irrelevant and low relevant areas A and B. In this study self-management with the CoaguChek XS was clinically safe and reliable.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19404552

Source DB:  PubMed          Journal:  Thromb Haemost        ISSN: 0340-6245            Impact factor:   5.249


  6 in total

1.  The need for clinical accuracy guidelines for blood glucose monitors.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Home management of INR in the public health system: feasibility of self-management of oral anticoagulation and long-term performance of individual POC devices in determining INR.

Authors:  Sabrina da Silva Saraiva; Fernanda Andrade Orsi; Marcos Pereira Santos; Tania Machado; Silmara Montalvão; Carolina Costa-Lima; Erich Vinícius de Paula; Marina Pereira Colella; Joyce Annichino-Bizzacchi
Journal:  J Thromb Thrombolysis       Date:  2016-07       Impact factor: 2.300

3.  Performance of point-of-care international normalized ratio measurement to diagnose trauma-induced coagulopathy.

Authors:  Thomas Mistral; Yvonnick Boué; Jean-Luc Bosson; Pauline Manhes; Jules Greze; Julien Brun; Pierre Albaladejo; Jean-François Payen; Pierre Bouzat
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2017-06-21       Impact factor: 2.953

4.  Comparison of prothrombin time derived from CoaguChek XS and laboratory test according to fibrinogen level.

Authors:  Sue Jung Kim; Eun Young Lee; Rojin Park; Juwon Kim; Jaewoo Song
Journal:  J Clin Lab Anal       Date:  2014-03-28       Impact factor: 2.352

5.  International normalized ratio testing with point-of-care coagulometer in healthy term neonates.

Authors:  Shigeo Iijima; Toru Baba; Daizo Ueno; Akira Ohishi
Journal:  BMC Pediatr       Date:  2014-07-09       Impact factor: 2.125

6.  Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic.

Authors:  Bushra Moiz; Anila Rashid; Muhammad Hasan; Lena Jafri; Ahmed Raheem
Journal:  Clin Appl Thromb Hemost       Date:  2018-01-28       Impact factor: 2.389

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.