Literature DB >> 19404117

Isolation of the brain-related factor of the error between intended and achieved position of deep brain stimulation electrodes implanted into the subthalamic nucleus for the treatment of Parkinson's disease.

Slawomir Daniluk1, Keith G Davies, Peter Novak, Thai Vu, Jules M Nazzaro, Samuel A Ellias.   

Abstract

OBJECTIVE: Although a few studies have quantified errors in the implantation of deep brain stimulation electrodes into the subthalamic nucleus (STN), a significant trend in error direction has not been reported. We have previously found that an error in axial plane, which is of most concern because it cannot be compensated for during deep brain stimulation programming, had a posteromedial trend. We hypothesized that this trend results from a predominance of a directionally oriented error factor of brain origin. Accordingly, elimination of nonbrain (technical) error factors could augment this trend. Thus, implantation accuracy could be improved by anterolateral compensation during target planning.
METHODS: Surgical technique was revised to minimize technical error factors. During 22 implantations, targets were selected on axial magnetic resonance imaging scans up to 1.5 mm anterolateral from the STN center. Using fusion of postoperative computed tomographic and preoperative magnetic resonance imaging scans, implantation errors in the axial plane were obtained and compared with distances from the lead to the STN to evaluate the benefit of anterolateral compensation.
RESULTS: Twenty errors and the mean error had a posteromedial direction. The average distances from the lead to the target and to the STN were 1.7 mm (range, 0.8-3.1 mm) and 1.1 mm (range, 0.1-1.9 mm), respectively. The difference between the 2 distances was significant (paired t test, P < 0.0001). The lower parts of the lead were consistently bent in the posteromedial direction on postoperative scout computed tomographic scans, suggesting that a brain-related factor is responsible for the reported error.
CONCLUSION: Elimination of the technical factors of error during STN deep brain stimulation implantation can result in a consistent posteromedial error. Implantation accuracy may be improved by compensation for this error in advance.

Entities:  

Mesh:

Year:  2009        PMID: 19404117     DOI: 10.1227/01.NEU.0000335171.38334.39

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  2 in total

1.  The importance of testing deep brain stimulation lead impedances before final lead implantation.

Authors:  Jules M Nazzaro; Kelly E Lyons; Rajesh Pahwa; Larry W Ridings
Journal:  Surg Neurol Int       Date:  2011-09-27

2.  Accuracy of Intraoperative Computed Tomography in Deep Brain Stimulation-A Prospective Noninferiority Study.

Authors:  Naomi I Kremer; D L Marinus Oterdoom; Peter Jan van Laar; Dan Piña-Fuentes; Teus van Laar; Gea Drost; Arjen L J van Hulzen; J Marc C van Dijk
Journal:  Neuromodulation       Date:  2019-01-10
  2 in total