Performance measurement of the scintillator with optical fiber detector for boron neutron capture therapy.

M Komeda1, H Kumada, M Ishikawa, T Nakamura, K Yamamoto, A Matsumura.   

Abstract

The thermal neutron flux can be easily measured in real time by using the scintillator with optical fiber (SOF) detector. However the irradiation damage under high-intensity neutron flux causes deterioration of the SOF detector due to radiation damage to the plastic scintillator in which (6)LiF is blended. After irradiating the SOF detector for 4 h (thermal neutron fluence is approximately 2.0 x 10(13)neutrons/cm(2)), the sensitivity of the SOF detector decreased by 3.0%. After irradiating the SOF detector for 2 months (thermal neutron fluence approximately 6.4 x 10(14)neutrons/cm(2)), the sensitivity was reduced to 42% of baseline. Supposing that the thermal neutron fluence is 2 x 10(12)neutrons/cm(2) on the surface of a patient in a BNCT treatment, the sensitivity of the SOF detector is reduced by approximately 0.3%. This report presents investigations on the deterioration of the SOF detector in irradiation experiments.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19398347     DOI: 10.1016/j.apradiso.2009.03.083

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Determining a methodology of dosimetric quality assurance for commercially available accelerator-based boron neutron capture therapy system.

Authors:  Katsumi Hirose; Takahiro Kato; Takaomi Harada; Tomoaki Motoyanagi; Hiroki Tanaka; Akihiko Takeuchi; Ryohei Kato; Shinya Komori; Yuhei Yamazaki; Kazuhiro Arai; Noriyuki Kadoya; Mariko Sato; Yoshihiro Takai
Journal:  J Radiat Res       Date:  2022-07-19       Impact factor: 2.438

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.