Literature DB >> 19396787

Quantitative assessment of anterior chamber volume using slit-lamp OCT and Pentacam.

Umut Dinc1, Banu Oncel, Ebru Gorgun, Levent Alimgil.   

Abstract

PURPOSE: To quantitatively assess and compare the anterior chamber volume (ACV) by different anterior segment imaging techniques.
METHODS: Forty healthy eyes of 40 patients were recruited and three consecutive measurements of ACV were determined prospectively utilizing SL-OCT and Pentacam. Statistical significance of interdevice differences between measurements was evaluated by unpaired t test and Bland-Altman analysis. The repeatability of three consecutive measurements was analyzed by repeated measured analysis of variance.
RESULTS: Eighty eyes of 40 patients (18 female, 22 male) with a mean spherical error of -0.55+/-1.4 diopters were included in the study. Mean ACV were detected as 163.34+/-29.76 mm3 by Pentacam and 164.14+/-28.44 mm3 by SL-OCT (unpaired t test, p=0.872). Pentacam and SL-OCT displayed a high intrasession repeatability (repeated measured analysis of variance, p>0.05). ACV values detected by SL-OCT were clinically interchangeable with Pentacam.
CONCLUSIONS: Noncontact measurements of ACV using SL-OCT and Pentacam were easy to handle and demonstrated good repeatability. The tested devices were regarded as compatible in measuring ACV.

Entities:  

Mesh:

Year:  2009        PMID: 19396787     DOI: 10.1177/112067210901900314

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  4 in total

1.  Comparison of the anterior ocular segment measurements using swept-source optical coherent tomography and a scanning peripheral anterior chamber depth analyzer.

Authors:  Toshie Furuya; Fumihiko Mabuchi; Tatsuya Chiba; Satoshi Kogure; Shigeo Tsukahara; Kenji Kashiwagi
Journal:  Jpn J Ophthalmol       Date:  2011-08-10       Impact factor: 2.447

2.  The role of Scheimpflug imaging in the management of posterior scleritis.

Authors:  Natalia Pawlowska; Jonathan Luck
Journal:  Eye Brain       Date:  2010-05-18

3.  Measurement of and Factors Associated with the Anterior Chamber Volume in Healthy Chinese Adults.

Authors:  Yuan Zong; Qian Xu; Chunhui Jiang; Haohao Zhu; Jian Yu; Xinghuai Sun
Journal:  J Ophthalmol       Date:  2017-01-10       Impact factor: 1.909

4.  Limitations in imaging common conjunctival and corneal pathologies with fourier-domain optical coherence tomography.

Authors:  Hakan Demirci; Daniel W Steen
Journal:  Middle East Afr J Ophthalmol       Date:  2014 Jul-Sep
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.