P Zuccaro1, I Altieri, P Betto, R Pacifici, G Ricciarello, L A Pini, E Sternieri, S Pichini.
Abstract
A simple and reliable reversed-phase high-performance liquid chromatographic method with coulometric detection is described for the quantitation of naltrexone and its metabolite, 6 beta-naltrexol, in plasma samples of healthy volunteers who received orally 50 mg of naltrexone. The analytes and the internal standard, naloxone, are extracted with an octadecyl solid-phase extraction column before chromatography. The mobile phase is 0.01 M potassium phosphate (pH 3)-acetonitrile (85:15, v/v) and it is pumped at 0.8 ml/min. The coulometric detector is formed by two electrodes set at +0.20 V and +0.70 V, with a palladium reference electrode. The limit of quantitation observed was 5 ng/ml for both naltrexone and 6 beta-naltrexol. This method can be used to investigate pharmacokinetic parameters of different pharmaceutical preparations of this opioid antagonist.Entities:
Mesh:
Substances:
Year: 1991 PMID: 1939481 DOI: 10.1016/0378-4347(91)80156-7
Source DB: PubMed Journal: J Chromatogr