Evaluation and advantages of an automatic magnetic mixing of syringes integrated to a whole blood gas analyser.

Marie-Odile Benoit1, Jean-Louis Paul.   

Abstract

BACKGROUND: Arterial blood gases are essential for the diagnostic and therapeutic management of severely ill patients. The pre-analytic phase is crucial to the quality and interpretation of the results. Whole blood specimens used for blood gas analysis must be thoroughly mixed prior to analysis. Producing a homogeneous sample is important to obtain accurate results. The availability of a new system able to automate the standardization of the sample mixing would be a way to improve this step. We evaluated the quality of automatic mixing with the use of SafePICO syringes and ABL 825 FLEX blood-gas analyser.
METHODS: Quantitative measurements of total hemoglobin and potassium were performed on ABL 825 FLEX from 216 whole blood specimens collected in SafePICO syringes and were compared with those obtained with LH 750 for hemoglobin and LX 20 for potassium.
RESULTS: Quantitative results showed excellent agreement with a reference analyser for the analytes with correlation coefficients greater than 0.97. No bias was observed for hemoglobin and the small difference observed for potassium is not clinically significant and was not due to hemolysis but to the differences between the methods.
CONCLUSIONS: ABL 825 FLEX produces a homogeneous specimen which minimizes the pre-analytical errors manipulator dependent.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19391061     DOI: 10.1080/00365510902914073

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  3 in total

1.  Evaluation of automatic mixing versus manual mixing for point of care hemoglobin measurement.

Authors:  Ghaith Altawallbeh; Pedro Castaneda; Gitte Wennecke; Amy B Karger
Journal:  Pract Lab Med       Date:  2020-04-05

2.  Role of electrolyte abnormalities and unmeasured anions in the metabolic acid-base abnormalities in dogs with parvoviral enteritis.

Authors:  Richard K Burchell; Arnon Gal; Ryan Friedlein; Andrew L Leisewitz
Journal:  J Vet Intern Med       Date:  2020-03-04       Impact factor: 3.333

3.  Preanalytical conditions of point-of-care testing in the intensive care unit are decisive for analysis reliability.

Authors:  Adrien Auvet; Fabien Espitalier; Leslie Grammatico-Guillon; Mai-Anh Nay; Djilali Elaroussi; Marc Laffon; Christian R Andres; Annick Legras; Stephan Ehrmann; Pierre-François Dequin; Chantal Gendrot; Antoine Guillon
Journal:  Ann Intensive Care       Date:  2016-06-24       Impact factor: 6.925

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.