Assessing the performance of the Whisperflow continuous positive airway pressure generator: a bench study.

G W Glover1, S J Fletcher.   

Abstract

BACKGROUND: There are few data describing the performance of the Whisperflow continuous positive airway pressure (CPAP) generator.
METHODS: (i) (a) A static test of 11 Whisperflow devices examining maximum flow generation with no load and with 2.5, 5, 7.5, 10, 15, and 20 cm H2O valve loading, at varying Fi(O2). (b) CPAP valves (Accu-peep(R), Vital Signs, Totowa, NJ, USA) were tested by measuring mean upstream pressure at varying flows in five valves (2.5, 5, 7.5, 10, 15, and 20 cm H2O). (ii) We measured the mean and minimum inspiratory mask pressure generated by a representative Whisperflow device in a model of spontaneous respiration. Measurements were made with combinations of Fi(O2), ventilatory frequency, tidal volume, and valve loading similar to those encountered in clinical practice.
RESULTS: (i) (a) The flow generated by the Whisperflow valves decreases with increasing valve load and increasing Fi(O2) (from 140 to 20 litre min(-1)). (b) The CPAP valves maintain the required pressure within acceptable limits against varying flow. (ii) At all permutations, the mean inspiratory mask pressure was significantly lower than that required. At high inspiratory flow rates, the minimum inspiratory pressure approached atmospheric pressure.
CONCLUSIONS: The Whisperflow may not perform as expected. Clinicians should be cautious when using this device, particularly with high Fi(O2) and CPAP valve load. The flow setting should be set at maximum. Failure of CPAP therapy may be due to failure of the generator. Further in vivo data are required.

Entities:  

Mesh:

Year:  2009        PMID: 19380311     DOI: 10.1093/bja/aep077

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  4 in total

1.  CPAP for acute cardiogenic pulmonary oedema from out-of-hospital to cardiac intensive care unit: a randomised multicentre study.

Authors:  Laurent Ducros; Damien Logeart; Eric Vicaut; Patrick Henry; Patrick Plaisance; Jean-Philippe Collet; Claire Broche; Papa Gueye; Muriel Vergne; David Goetgheber; Pierre-Yves Pennec; Vanessa Belpomme; Jean-Michel Tartière; Sophie Lagarde; Marius Placente; Marie-Laurence Fievet; Gilles Montalescot; Didier Payen
Journal:  Intensive Care Med       Date:  2011-07-30       Impact factor: 17.440

2.  Conceptual model of low-cost improvised bubble continuous positive airway pressure device for adults and its potential use in the COVID-19 pandemic.

Authors:  Himal Kharel; Zeni Kharel; Samikchhya Keshary Bhandari
Journal:  PLoS Negl Trop Dis       Date:  2022-03-03

3.  Performance of EasyBreath Decathlon Snorkeling mask for delivering continuous positive airway pressure.

Authors:  Alberto Noto; Claudia Crimi; Andrea Cortegiani; Massimiliano Giardina; Filippo Benedetto; Pietro Princi; Annalisa Carlucci; Lorenzo Appendini; Cesare Gregoretti
Journal:  Sci Rep       Date:  2021-03-10       Impact factor: 4.379

4.  The LeVe CPAP System for Oxygen-Efficient CPAP Respiratory Support: Development and Pilot Evaluation.

Authors:  Pete Culmer; W Davis Birch; I Waters; A Keeling; C Osnes; D Jones; G de Boer; R Hetherington; S Ashton; M Latham; T Beacon; T Royston; R Miller; A Littlejohns; J Parmar; Tom Lawton; S Murdoch; D Brettle; R Musasizi; G Nampiina; E Namulema; N Kapur
Journal:  Front Med Technol       Date:  2021-08-24
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.