Literature DB >> 19374303

Reliability of digital compass goniometer in knee joint range of motion measurement.

Nammond Yaikwawongs1, Noppachart Limpaphayom, Vajara Wilairatana.   

Abstract

OBJECTIVE: To compare the reliability of range of motion measurement in the knee joint using a digital compass goniometer combined with inclinometer with standard range of motion measurement from roentgenographic picture. MATERIAL AND
METHOD: Range of flexion and extension of the knee joint in volunteer participants was measured by the newly developed digital compass goniometer combined with inclinometer (DCG). The results were compared with range of knee joint motion obtained from standard roentgenographic picture by intraclass correlation coefficient.
RESULTS: Range of motion of knee joint measured by DCG correlated very well with the data obtained from standard knee roentgenographic picture. The intraclass correlation coefficient equals 0.973.
CONCLUSION: The digital compass goniometer was a reliable tool to measure knee joint range of motion in flexion and extension plane.

Entities:  

Mesh:

Year:  2009        PMID: 19374303

Source DB:  PubMed          Journal:  J Med Assoc Thai        ISSN: 0125-2208


  5 in total

1.  Measurement of the knee flexion angle with smartphone applications: Which technology is better?

Authors:  Jean-Yves Jenny; Abdullah Bureggah; Yann Diesinger
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-02-15       Impact factor: 4.342

2.  Assessment of Correlation between Electrogoniometer Measurements and Sports-Specific Movement in Karate Elites.

Authors:  Sara Lotfian; Afsaneh Safar Cherati; Aliashraf Jamshidi; Mohammad Ali Sanjari
Journal:  Asian J Sports Med       Date:  2014-06

3.  Physical activity after outpatient surgery and enhanced recovery for total knee arthroplasty.

Authors:  M G M Schotanus; Y F L Bemelmans; B Grimm; I C Heyligers; N P Kort
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-08-04       Impact factor: 4.342

4.  Successful management of total knee replacement in a high responder hemophilia patient with a history of inhibitor.

Authors:  Roya Dolatkhah; Mohammad Reza Bazavar; Masoud Poureisa; Iraj Asvadi Kermani; Jalil Vaez Gharamaleki; Zohreh Sanaat; Jamal Eivazi Ziaei; Alireza Nikanfar; Ali Esfahani; Seyed Hadi Chavoshi
Journal:  Iran Red Crescent Med J       Date:  2013-01-05       Impact factor: 0.611

5.  Smartphone and Universal Goniometer for Measurement of Elbow Joint Motions: A Comparative Study.

Authors:  Behnam Behnoush; Nasim Tavakoli; Elham Bazmi; Fariborz Nateghi Fard; Mohammad Hossein Pourgharib Shahi; Arash Okazi; Tahmineh Mokhtari
Journal:  Asian J Sports Med       Date:  2016-06-11
  5 in total