Disassembly of a modular femoral component after femoral head prosthetic replacement.

Toshiki Shiga1, Masaki Mori, Tatsuro Hayashida, Yasuhiro Fujiwara, Taku Ogura.   

Abstract

We report a rare case of a taper-locked femoral inner head that have been completely separated from the stem neck. A 76-year-old man who had hip fracture of the right hip had disassembly of inner head and stem neck after revision of bipolar hip prosthesis. This force could have been amplified by the pumping phenomenon generated after the revised inner head. We suspect the sealed air pushes back the stem neck, and unlocks the taper lock of the inner head, causing separation of the stem neck from the inner head. To prevent pumping phenomenon, we recommend manual testing of the taper lock to confirm that it has been assembled correctly and its integrity before implantation when the bloody and fatty membrane is adequately removed from stem neck. Copyright 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19361950     DOI: 10.1016/j.arth.2009.03.001

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Impending spontaneous head neck dissociation caused by anteverted cup: A case report on reverse bottle opener effect and review of literature.

Authors:  Ramneek Mahajan; Piyush Suresh Nashikkar; Varun Khanna
Journal:  J Clin Orthop Trauma       Date:  2020-11-01

Review 2.  How have alternative bearings and modularity affected revision rates in total hip arthroplasty?

Authors:  William M Mihalko; Markus A Wimmer; Carol A Pacione; Michel P Laurent; Robert F Murphy; Carson Rider
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 3.  Review: Trunnionosis leading to modular femoral head dissociation.

Authors:  Agneish Dutta; James Nutt; Guy Slater; Syed Ahmed
Journal:  J Orthop       Date:  2021-01-30

4.  Late Nontraumatic Dissociation of the Femoral Head and Trunnion in a Total Hip Arthroplasty.

Authors:  Simon J M Parker; Wasim Khan; Simon Mellor
Journal:  Case Rep Orthop       Date:  2015-05-11

5.  Spontaneous Dissociation of Anatomic Medullary Locking A Plus (AML A Plus) Femoral Component at the Head-Neck Interface.

Authors:  Ketan Pande; Juzaily Fekry Leong; Ngai Nung Lo
Journal:  J Orthop Case Rep       Date:  2015 Jul-Sep

6.  Spontaneous dissociation of bipolar hip hemiarthroplasty in a patient with nerve palsy: A case report and review of the literature.

Authors:  Varah Yuenyongviwat; Khanin Iamthanaporn; Theerawit Hongnaparak
Journal:  Int J Surg Case Rep       Date:  2015-08-29

7.  Late Onset Disassembly of a Modular Neck-stem Component after Cementless Hip Replacement without Dislocation: A Case Report and Review of Literatures.

Authors:  Ioannis Papaioannou; Thomas Repantis; Andreas Baikousis; Panagiotis Korovessis
Journal:  Hip Pelvis       Date:  2018-09-04

8.  Disassembly of the inner head of a bipolar hip prosthesis.

Authors:  Tsuyoshi Toyota; Hiroshi Horiuchi; Yoshiyuki Nakamura; Ikuya Yamazaki; Tsutomu Takizawa
Journal:  Arthroplast Today       Date:  2018-08-31
  8 in total