Literature DB >> 19360859

Experience with the Cardiva Boomerang Catalyst system in pediatric cardiac catheterization.

Sharon Seltzer1, Juan Carlos Alejos, Daniel S Levi.   

Abstract

OBJECTIVES: We studied the safety and efficacy of the Cardiva Boomerang Catalyst vascular closure system in pediatric patients after cardiac catheterization with access in femoral and internal jugular vessels.
BACKGROUND: Recurrent catheterization and advances in pediatric interventions increase the need for easy hemostasis without a residual foreign body that may prevent re-accessing the vessel. The Boomerang can be deployed in sheaths as small as 4Fr without residual foreign body, with minimal orientation needed, and few complications reported.
METHODS: In a two-month period, all patients between 18 months and 21 years old catheterized with 4-8Fr sheaths less than 15 cm long were eligible for Boomerang placement. These were compared retrospectively with control patients with manual hemostasis. Anthropomorphic measurements, procedure type, activated clotting time, and sheath size as well as total times of cases, intubation, hemostasis, and extubation were compared between the two groups.
RESULTS: Forty-six Boomerangs were deployed in 31 patients and compared with 40 patients with manual hemostasis. Boomerangs were deployed in femoral vessels and the internal jugular vein. Device success with hemostasis was achieved in 39 patients (85%). There were no significant differences in time to hemostasis or extubation between the two groups. No major complications or operator error occurred, including hematoma, transfusion, retroperitoneal bleed, infection, vessel occlusion, or need for surgery.
CONCLUSIONS: The Boomerang is a safe and easy means of achieving hemostasis in the pediatric population, in femoral vessels as well as internal jugular veins. Its times to hemostasis and extubation were not significantly different from manual hold. 2009 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2009        PMID: 19360859     DOI: 10.1002/ccd.22002

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

1.  Successful use of the Cardiva Boomerang™ vascular closure device to close a brachial artery puncture site after emergency PTCA.

Authors:  Plinio Cirillo; Gianluca Petrillo; Greta Luana D'Ascoli; Federico Piscione; Massimo Chiariello
Journal:  Heart Vessels       Date:  2010-09-28       Impact factor: 2.037

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.